Improvements in Remote Cardiopulmonary Measurement Using a Five Band Digital Camera
Name
Picard_Improvements in remote.pdf
Size
2.75 MB
Format
Adobe PDF
Checksum (MD5)
111cd451c74aff55f1ffaa5df58935ed
Author(s) • •
McDuff, Daniel Jonathan
Gontarek, Sarah
Picard, Rosalind W.
Date Issued
May 2014
Journal
IEEE Transactions on Biomedical Engineering
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
McDuff, Daniel; Gontarek, Sarah and Picard, Rosalind W. “Improvements in Remote Cardiopulmonary Measurement Using a Five Band Digital Camera.” IEEE Transactions on Biomedical Engineering 61, no. 10 (October 2014): 2593–2601. © 2014 Institute of Electrical and Electronics Engineers (IEEE)
Version
Author's final manuscript
Abstract
Remote measurement of the blood volume pulse via photoplethysmography (PPG) using digital cameras and ambient light has great potential for healthcare and affective computing. However, traditional RGB cameras have limited frequency resolution. We present results of PPG measurements from a novel five band camera and show that alternate frequency bands, in particular an orange band, allowed physiological measurements much more highly correlated with an FDA approved contact PPG sensor. In a study with participants (n = 10) at rest and under stress, correlations of over 0.92 (p <; 0.01) were obtained for heart rate, breathing rate, and heart rate variability measurements. In addition, the remotely measured heart rate variability spectrograms closely matched those from the contact approach. The best results were obtained using a combination of cyan, green, and orange (CGO) bands; incorporating red and blue channel observations did not improve performance. In short, RGB is not optimal for this problem: CGO is better. Incorporating alternative color channel sensors should not increase the cost of such cameras dramatically.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Media Laboratory
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1109/TBME.2014.2323695