Towards Attogram Mass Measurements in Solution with Suspended Nanochannel Resonators
Name
Manalis-Toward Attogram Mass Measurements in Solution with Suspended Nanochannel Resonators.pdf
Size
1.02 MB
Format
Adobe PDF
Checksum (MD5)
bfab012964369a548e51bbf702f32207
Author(s) • • • •
Lee, Jungchul
Shen, Wenjiang
Payer, Kristofor Robert
Burg, Thomas P.
Manalis, Scott R.
Alternative Title
Toward Attogram Mass Measurements in Solution with Suspended Nanochannel Resonators
Date Issued
July 2011
Journal
Nano Letters
Publisher
American Chemical Society
Citation
Lee, Jungchul et al. “Toward Attogram Mass Measurements in Solution with Suspended Nanochannel Resonators.” Nano Letters 10.7 (2010): 2537-2542.
Version
Author's final manuscript
Abstract
Using suspended nanochannel resonators (SNRs) we demonstrate measurements of mass in solution with a resolution of 27 ag in a 1 kHz bandwidth, which represents a 100-fold improvement over existing suspended microchannel resonators and, to our knowledge, is the most precise mass measurement in liquid today. The SNR consists of a cantilever that is 50 μm long, 10 μm wide, and 1.3 μm thick, with an embedded nanochannel that is 2 μm wide and 700 nm tall. The SNR has a resonance frequency near 630 kHz and exhibits a quality factor of approximately 8,000 when dry and when filled with water. In addition, we introduce a new method that uses centrifugal force caused by vibration of the cantilever to trap particles at the free end. This approach eliminates the intrinsic position dependent error of the SNR and also improves the mass resolution by increasing the averaging time for each particle.
MIT Department
Massachusetts Institute of Technology. Department of Biological Engineering
Massachusetts Institute of Technology. Department of Mechanical Engineering
Massachusetts Institute of Technology. Microsystems Technology Laboratories
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike 3.0
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1021/nl101107u