Design of a prototype handheld nanoliter pipette
Name
964449970-MIT.pdf
Description
Full printable version
Size
5.88 MB
Format
Adobe PDF
Checksum (MD5)
e034c4b6e64a0161e13f90fc93f08ca8
Author(s)
Samsel, Adrian Christopher
Advisor(s)
Anastasios John Hart.
Date Issued
2016
Publisher
Massachusetts Institute of Technology
Abstract
Handheld pipettes are invaluable to scientists in labs across the globe. Due to high inaccuracies at volumes of less than 1 [mu]L, a nanoliter pipette could truly change the way people research. This experimental nanoliter pipette uses a regulating CAM mechanism and a diaphragm to displace small amounts of fluid. The CAM mechanism has the ability to be adjusted to different steps, each of which controls the amount of working fluid to be displaced in a sealed chamber. In response to this, the diaphragm displaces a smaller volumetric amount of fluid outside the chamber. A deamplification ratio is defined by the ratio of the volume displaced by the diaphragm to the volume displaced by the CAM mechanism. The deamplification ratio is adjustable by changing the setting of the variable CAM mechanism or exchanging different tip assemblies which hold different sized diaphragms. In all, this nanoliter pipette enables measuring and dispensing of very small volumes of liquid, ranging from I nL to 1[mu] L, and can expand the tools of every lab researcher internationally.
Description
Thesis: S.B., Massachusetts Institute of Technology, Department of Mechanical Engineering, 2016.
Cataloged from PDF version of thesis.
Includes bibliographical references (page 26).
Subjects
Mechanical Engineering.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Persistent DSpace Link