Optimal primary coil size for wireless power telemetry to medical implants
Name
Wyatt_Optimal primary.pdf
Size
227.09 KB
Format
Adobe PDF
Checksum (MD5)
78b3765416c2b47e210879ebcc709b5a
Author(s) • • • •
Wyatt, John L.
Mendoza, Oscar D.
Doyle, Patrick S.
Kelly, Shawn K.
Priplata, Attila
Date Issued
January 2011
Journal
Proceedings of the IEEE International Symposium on Applied Sciences in Biomedical and Communication Technologies (ISABEL), 2010
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Wyatt, John L. et al. "Optimal primary coil size for wireless power telemetry to medical implants." Proceedings of the IEEE International Symposium on Applied Sciences in Biomedical and Communication Technologies (ISABEL), 2010: 1-5. © IEEE 2010.
Version
Final published version
Abstract
A retinal prosthesis telemetry system is examined, and several methods are explored to optimize the size of the external primary telemetry coil to maximize the wireless delivery of power to an implanted secondary coil of constrained size. A simplified version of the Biot-Savart Law is used to give a first-pass optimal primary coil size for a small secondary coil. Numerical integration is then used to improve the optimization for larger secondary coils, and this calculation is repeated across a range of secondary coil radii. Finally, the effects of eye rotation angle are explored, with the future goal of expanding the optimization techniques to cover the predicted range of angular eye excursions.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1109/ISABEL.2010.5702795