Drag reducing polymers as simple indicators of hemolytic potential in biomechanical devices
Name
558634750-MIT.pdf
Description
Full printable version
Size
5.56 MB
Format
Adobe PDF
Checksum (MD5)
c2409a63307080ca7273989d316d42a7
Author(s)
Shieh, Sarah
Advisor(s)
Anette E. Hosoi.
Date Issued
2009
Publisher
Massachusetts Institute of Technology
Abstract
An experimental study was carried out to determine if drag reducing polymers can be simple indicators of hemolytic potential in biomechanical devices. Specifically, three different blood pumps, known as a left ventricle assist devices (LVADs) were operated in a test loop using an aqueous solution of polyethylene glycol (PEO, MW = 5000 kDa), a known drag reducing polymer. The pumps were operated under controlled parameters and the change in viscosity (cP) and drag reduction (%DR) for each pump was monitored over the specified time period. The CentriMag® (CM) was used to confirm the drag reducing behavior of PEO, while HeartMate® II (HM II) and HeartMate® III (HM III) were used to determine if there was a correlation between experimental results and actual hemolysis results. Experimental results showed that the mathematical difference between the average final and initial viscosity of HM II was greater than the difference for HM III. HM II had a difference of 0.21 cP and HM III had a difference of 0.16 cP. Hemolysis results using bovine blood showed that HM II had a higher hemolysis rate of 3.80 +/- 1.11 g/day and a higher milligram normalized index of hemolysis of 0.0393 +/- 0.0155. The average hemolysis rate for HM III was 1.38 +/- 0.63 g/day and the milligram normalized index of hemolysis (mg N.I.H.) was 0.571 +/- 0.333. This positive correlation shows that PEO can be a simple indicator of hemolytic potential for biomechanical devices. More data and experimentation is needed to further understand the behavior of PEO and it's ability to indicate hemolytic potential using a wider range of biomechanical devices.
Description
Thesis (S.B.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2009.
Cataloged from PDF version of thesis.
Includes bibliographical references (p. 23-24).
Subjects
Mechanical Engineering.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
M.I.T. theses are protected by
copyright. They may be viewed from this source for any purpose, but
reproduction or distribution in any format is prohibited without written
permission. See provided URL for inquiries about permission.
copyright. They may be viewed from this source for any purpose, but
reproduction or distribution in any format is prohibited without written
permission. See provided URL for inquiries about permission.
Persistent DSpace Link