Simulated radiation levels and patterns of MRI without a Faraday shielded room

Thumbnail Image
Download
Name

Magnetic Resonance in Med - 2025 - Kazemivalipour - Simulated radiation levels and patterns of MRI without a Faraday.pdf

Size

3.82 MB

Format

Adobe PDF

Checksum (MD5)

365909173ca6de78f828fea8bf2435af

Author(s)
Kazemivalipour, Ehsan
•
Guerin, Bastien
•
Wald, Lawrence L.
Date Issued
March 17, 2025
Journal
Magnetic Resonance in Medicine
Publisher
Wiley
Citation
Kazemivalipour E, Guerin B, Wald LL. Simulated radiation levels and patterns of MRI without a Faraday shielded room. Magn Reson Med. 2025; 94: 835-851.
Version
Final published version
Abstract
Purpose: We characterize electromagnetic (EM) radiation patterns and levelsin conventional MRI systems as a function of field strength and load symmetry,providing a framework for mitigation strategies allowing operation without ashielded room. Methods: We simulated the far-field radiation pattern and fields at a 10 mradius (|E|10m and |B|10m ) for a solenoidal superconducting MRI with abody birdcage coil operated between 0.25T and 6.5T. Five load configura-tions probed the impact of load-symmetry, ranging from a sphere to a bodyload (least-symmetric). We also assessed simple layered EM absorbers at thebore-ends. Results: All configurations exceeded regulatory limits for realistic transmit lev-els. At 1.5T, a 300 V rms RF-pulse is 2700-fold the |E|10m limit. Field strengthand load symmetry strongly modulate radiation patterns and levels. The radi-ated power increased by more than four orders of magnitude from 0.25T to6.5T. Spherical load radiation transitioned from a peak gain at the bore-ends(0.25–0.5T) to a donut-shaped pattern, suggesting current loops around the bore(1 T–1.5T), back to bore-axis-directed gain, suggesting propagating waves alongthe bore (2T–6.5T). Transition patterns were seen between these regimes; uni-form radiation at 0.75T and a combined donut/bore-directed pattern at 1.75T.Load asymmetry increased both strength and pattern asymmetry, with the bodyload having the highest and least symmetric radiation with the legs facilitat-ing wave propagation at high-fields. A simple optimized layered absorber atscanner’s service-end reduced 3T peak radiation by 11 dB. Conclusion: Radiation from unshielded scanners far exceeds regulatory lim-its, particularly at high-field. Mitigation strategies must address load-symmetry,field strength, and wave effects.
MIT Department
Harvard-MIT Program in Health Sciences and Technology
Terms of Use
Creative Commons Attribution
https://creativecommons.org/licenses/by/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/162855
DOI of Published Version
https://doi.org/10.1002/mrm.30499
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.