Using collision-induced dissociation to constrain sensitivity of ammonia chemical ionization mass spectrometry (NH 4 + CIMS) to oxygenated volatile organic compounds

Thumbnail Image
Download
Name

amt-12-1861-2019.pdf

Description
Published version
Size

378.98 KB

Format

Adobe PDF

Checksum (MD5)

f11407ea5660f7e0b04b58f723618e46

sword-2020-05-11T18:59:23.original.xml (130 B)
Original SWORD entry document
Author(s)
Koss, Abigail R.
•
Lim, Christopher Yung-Ta
•
Rowe, James Clifford.
•
Kroll, Jesse
Date Issued
March 2019
Journal
Atmospheric Measurement Techniques
Publisher
Copernicus GmbH
Citation
Zaytsev, Alexander et al. “Using collision-induced dissociation to constrain sensitivity of ammonia chemical ionization mass spectrometry (NH 4 + CIMS) to oxygenated volatile organic compounds” Atmospheric Measurement Techniques, vol. 12, no. 3, 2019, pp. 1861-1870 © 2019 The Author(s)
Version
Final published version
Abstract
Chemical ionization mass spectrometry (CIMS) instruments routinely detect hundreds of oxidized organic compounds in the atmosphere. A major limitation of these instruments is the uncertainty in their sensitivity to many of the detected ions. We describe the development of a new highresolution time-of-flight chemical ionization mass spectrometer that operates in one of two ionization modes: Using either ammonium ion ligand-switching reactions such as for NHC 4 CIMS or proton transfer reactions such as for protontransfer-reaction mass spectrometer (PTR-MS). Switching between the modes can be done within 2 min. The NH+4 CIMS mode of the new instrument has sensitivities of up to 67 000 dcps ppbv..1 (duty-cycle-corrected ion counts per second per part per billion by volume) and detection limits between 1 and 60 pptv at 2σ for a 1 s integration time for numerous oxygenated volatile organic compounds. We present a mass spectrometric voltage scanning procedure based on collision-induced dissociation that allows us to determine the stability of ammonium-organic ions detected by the NH+4 CIMS instrument. Using this procedure, we can effectively constrain the sensitivity of the ammonia chemical ionization mass spectrometer to a wide range of detected oxidized volatile organic compounds for which no calibration standards exist. We demonstrate the application of this procedure by quantifying the composition of secondary organic aerosols in a series of laboratory experiments.
MIT Department
Massachusetts Institute of Technology. Department of Civil and Environmental Engineering
Terms of Use
Creative Commons Attribution 4.0 International license
https://creativecommons.org/licenses/by/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/125653
DOI of Published Version
https://doi.org/10.5194/AMT-12-1861-2019
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.