  4. Characterization of the image plate multi-scan response to mono-energetic x-rays

Characterization of the image plate multi-scan response to mono-energetic x-rays

Thumbnail Image
Download
Name

093535_1_5.0219509 (1).pdf

Description
Published version
Size

4.42 MB

Format

Adobe PDF

Checksum (MD5)

081f4425ff512b5b9fd83af0bceaa058

sword-2026-03-10T14:43:49.original.xml (130 B)
Original SWORD entry document
Author(s)
Cufari, M
•
Vanderloo, N
•
Buschmann, BI
•
DeVault, A
•
Foo, BC
•
Vargas, J
•
Dannhoff, SG
•
Evans, TE
•
Johnson, TM
•
Kunimune, J
more
Date Issued
September 24, 2024
Journal
Review of Scientific Instruments
Publisher
AIP Publishing
Citation
M. Cufari, N. Vanderloo, B. I. Buschmann, A. DeVault, B. C. Foo, J. Vargas, S. G. Dannhoff, T. E. Evans, T. M. Johnson, J. Kunimune, Y. Lawrence, J. A. Pearcy, B. L. Reichelt, L. Russell, C. W. Wink, M. Gatu Johnson, R. D. Petrasso, J. A. Frenje; Characterization of the image plate multi-scan response to mono-energetic x-rays. Rev. Sci. Instrum. 1 September 2024; 95 (9): 093535.
Version
Final published version
Abstract
Image plates (IPs), or phosphor storage screens, are a technology employed frequently in inertial confinement fusion (ICF) and high energy density plasma (HEDP) diagnostics because of their sensitivity to many types of radiation, including, x rays, protons, alphas, beta particles, and neutrons. Prior studies characterizing IPs are predicated on the signal level remaining below the scanner saturation threshold. Since the scanning process removes some signal from the IP via photostimulated luminescence, repeatedly scanning an IP can bring the signal level below the scanner saturation threshold. This process, in turn, raises concerns about the signal response of IPs after an arbitrary number of scans and whether such a process yields, for example, a constant ratio of signal between the nth and n + 1st scan. Here, the sensitivity of IPs is investigated when scanned multiple times. It is demonstrated that the ratio of signal decay is not a constant with the number of scans and that the signal decay depends on the x-ray energy. As such, repeatedly scanning an IP with a mixture of signal types (e.g., x ray, neutron, and protons) enables ICF and HEDP diagnostics employing IPs to better isolate a particular signal type.
MIT Department
Massachusetts Institute of Technology. Plasma Science and Fusion Center
Terms of Use
Creative Commons Attribution
https://creativecommons.org/licenses/by/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/165058
DOI of Published Version
https://doi.org/10.1063/5.0219509
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.