Active fume hood sash height monitoring with audible feedback
Name
1-s2.0-S2352484718302634-main.pdf
Description
Published version
Size
1.38 MB
Format
Adobe PDF
Checksum (MD5)
34e8821ec523b21963eb466697f1c188
Author(s) • • • • • •
Becerra, Laura L
Ferrua, Juan A
Drake, Maxwell J
Kumar, Dheekshita
Anders, Ariel S
Wang, Evelyn N
Preston, Daniel J
Date Issued
2018
Journal
Energy Reports
Publisher
Elsevier BV
Version
Final published version
Abstract
© 2018 The Authors Fume hoods in laboratories consume the energy equivalent of up to four American households per hood; however, closing a modern hood's sash completely can save up to 75% of that energy. Past efforts have attempted to harness this potential energy reduction by reminding users to close the sash when a hood is not in use. In this work, we developed a device to measure the efficacy of these energy-saving methods. The device records the position of the sash and detects motion to determine whether a user is present, and, when fitted with a piezoelectric buzzer, can audibly alert users to close the sash when not in use. We installed this device in laboratories to quantify the energy and cost savings resulting from real-time audible feedback and found that the alarm reduced wasted energy by 87 to 98%. In addition, the platform demonstrated here can be used to quantitatively test other energy-saving methods that rely on user behavioral change in future work.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
Creative Commons Attribution 4.0 International license
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1016/J.EGYR.2018.09.010