In situ measurement of localization error in particle tracking microrheology
Name
397_2018_1117_ReferencePDF.pdf
Size
335.07 KB
Format
Adobe PDF
Checksum (MD5)
5dc7474798be0c4e72f0d4c9f6ea37aa
Author(s) • •
Josephson, Lilian L
Swan, James W
Furst, Eric M
Date Issued
October 27, 2018
Publisher
Springer Berlin Heidelberg
Version
Author's final manuscript
Abstract
Abstract
A method is presented to estimate the static error 𝜖 in a multiple particle tracking microrheology experiment. This in situ estimate of 𝜖 is measured under the same conditions of the material under test, and without any additional experiments. The correction of the mean-squared displacement by the in situ method is potentially more reliable than methods that rely on characterizing 𝜖 in a separate gel sample. With the true mean-squared displacements accessible at short lag times, experimental artifacts introduced by static error can be distinguished from true rheological properties, even in highly viscous (>10,000 mPa⋅s) samples.