Non-rigid registration of 3D ultrasound for neurosurgery using automatic feature detection and matching
Name
nihms973692.pdf
Description
Accepted version
Size
1.26 MB
Format
Adobe PDF
Checksum (MD5)
5f7d5a4f9fc3282847feecadfd82b564
Author(s) •
Golland, Polina
Wells, William M.
Date Issued
October 2018
Journal
International Journal of Computer Assisted Radiology and Surgery
Publisher
Springer Nature
Citation
Machado, Inês et al. “Non-rigid registration of 3D ultrasound for neurosurgery using automatic feature detection and matching.” International Journal of Computer Assisted Radiology and Surgery, vol. 13, no. 10, 2018, pp. 1525-1538 © 2018 The Author(s)
Version
Author's final manuscript
Abstract
Purpose: The brain undergoes significant structural change over the course of neurosurgery, including highly nonlinear deformation and resection. It can be informative to recover the spatial mapping between structures identified in preoperative surgical planning and the intraoperative state of the brain. We present a novel feature-based method for achieving robust, fully automatic deformable registration of intraoperative neurosurgical ultrasound images. Methods: A sparse set of local image feature correspondences is first estimated between ultrasound image pairs, after which rigid, affine and thin-plate spline models are used to estimate dense mappings throughout the image. Correspondences are derived from 3D features, distinctive generic image patterns that are automatically extracted from 3D ultrasound images and characterized in terms of their geometry (i.e., location, scale, and orientation) and a descriptor of local image appearance. Feature correspondences between ultrasound images are achieved based on a nearest-neighbor descriptor matching and probabilistic voting model similar to the Hough transform. Results: Experiments demonstrate our method on intraoperative ultrasound images acquired before and after opening of the dura mater, during resection and after resection in nine clinical cases. A total of 1620 automatically extracted 3D feature correspondences were manually validated by eleven experts and used to guide the registration. Then, using manually labeled corresponding landmarks in the pre- and post-resection ultrasound images, we show that our feature-based registration reduces the mean target registration error from an initial value of 3.3 to 1.5 mm. Conclusions: This result demonstrates that the 3D features promise to offer a robust and accurate solution for 3D ultrasound registration and to correct for brain shift in image-guided neurosurgery.
MIT Department
Massachusetts Institute of Technology. Computer Science and Artificial Intelligence Laboratory
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1007/S11548-018-1786-7