  4. Choroidal Neovascularization Analyzed on Ultrahigh-Speed Swept-Source Optical Coherence Tomography Angiography Compared to Spectral-Domain Optical Coherence Tomography Angiography

Choroidal Neovascularization Analyzed on Ultrahigh-Speed Swept-Source Optical Coherence Tomography Angiography Compared to Spectral-Domain Optical Coherence Tomography Angiography

Thumbnail Image
Download
Name

Choroidal neovascularization.pdf

Size

760.07 KB

Format

Adobe PDF

Checksum (MD5)

43b60172adabba755fb0898c6501a2a5

Author(s)
Novais, Eduardo A.
•
Louzada, Ricardo N.
•
Dang, Sabin
•
Regatieri, Caio V.S.
•
Witkin, André J.
•
Baumal, Caroline R.
•
Hornegger, Joachim
•
Jayaraman, Vijaysekhar
•
Duker, Jay S.
•
Waheed, Nadia K.
more
Date Issued
February 2016
Journal
American Journal of Ophthalmology
Publisher
Elsevier
Citation
Novais, Eduardo A.; Adhi, Mehreen; Moult, Eric M. et al. “Choroidal Neovascularization Analyzed on Ultrahigh-Speed Swept-Source Optical Coherence Tomography Angiography Compared to Spectral-Domain Optical Coherence Tomography Angiography.” American Journal of Ophthalmology 164 (April 2016): 80–88 © 2016 Elsevier Inc
Version
Author's final manuscript
Abstract
Purpose To compare visualization of choroidal neovascularization (CNV) secondary to age-related macular degeneration (AMD) using an ultrahigh-speed swept-source (SS) optical coherence tomography angiography (OCTA) prototype vs a spectral-domain (SD) OCTA device. Design Comparative analysis of diagnostic instruments. Methods Patients were prospectively recruited to be imaged on SD OCT and SS OCT devices on the same day. The SD OCT device employed is the RTVue Avanti (Optovue, Inc, Fremont, California, USA), which operates at ∼840 nm wavelength and 70 000 A-scans/second. The SS OCT device used is an ultrahigh-speed long-wavelength prototype that operates at ∼1050 nm wavelength and 400 000 A-scans/second. Two observers independently measured the CNV area on OCTA en face images from the 2 devices. The nonparametric Wilcoxon signed rank test was used to compare area measurements and P values of <.05 were considered statistically significant. Results Fourteen eyes from 13 patients were enrolled. The CNV in 11 eyes (78.6%) were classified as type 1, 2 eyes (14.3%) as type 2, and 1 eye (7.1%) as mixed type. Total CNV area measured using SS OCT and SD OCT 3 mm × 3 mm OCTA were 0.949 ± 1.168 mm² and 0.340 ± 0.301 mm², respectively (P = .001). For the 6 mm × 6 mm OCTA the total CNV area using SS OCT and SD OCT were 1.218 ± 1.284 mm² and 0.604 ± 0.597 mm², respectively (P = .0019). The field of view did not significantly affect the measured CNV area (P = .19 and P = .18 for SS OCT and SD OCT, respectively). Conclusion SS OCTA yielded significantly larger CNV areas than SD OCTA. It is possible that SS OCTA is better able to demarcate the full extent of CNV vasculature.
MIT Department
Massachusetts Institute of Technology. Institute for Medical Engineering & Science
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Research Laboratory of Electronics
Terms of Use
Creative Commons Attribution-NonCommercial-NoDerivs License
http://creativecommons.org/licenses/by-nc-nd/4.0/
Persistent DSpace Link
http://hdl.handle.net/1721.1/110907
DOI of Published Version
https://doi.org/10.1016/j.ajo.2016.01.011
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.