A multimodal spectroscopy system for real-time disease diagnosis
Name
Scepanovic_2009.pdf
Size
585.65 KB
Format
Adobe PDF
Checksum (MD5)
a93dae0d472d0af2039b63788ccaba9f
Author(s) • • • • •
Scepanovic, Obrad R.
Volynskaya, Zoya I.
Kong, Chae-Ryon
Galindo, Luis H.
Dasari, Ramachandra Rao
Feld, Michael S.
Date Issued
April 2009
Journal
Review of Scientific Instruments
Publisher
American Institute of Physics
Citation
Scepanovic, Obrad R. et al. “A multimodal spectroscopy system for real-time disease diagnosis.” Review of Scientific Instruments 80.4 (2009): 043103-9. © 2009 American Institute of Physics
Version
Final published version
Abstract
The combination of reflectance, fluorescence, and Raman spectroscopy—termed multimodal spectroscopy (MMS)—provides complementary and depth-sensitive information about tissue composition. As such, MMS is a promising tool for disease diagnosis, particularly in atherosclerosis and breast cancer. We have developed an integrated MMS instrument and optical fiber spectral probe for simultaneous collection of all three modalities in a clinical setting. The MMS instrument multiplexes three excitation sources, a xenon flash lamp (370–740 nm), a nitrogen laser (337 nm), and a diode laser (830 nm), through the MMS probe to excite tissue and collect the spectra. The spectra are recorded on two spectrograph/charge-coupled device modules, one optimized for visible wavelengths (reflectance and fluorescence) and the other for the near-infrared (Raman), and processed to provide diagnostic parameters. We also describe the design and calibration of a unitary MMS optical fiber probe 2 mm in outer diameter, containing a single appropriately filtered excitation fiber and a ring of 15 collection fibers, with separate groups of appropriately filtered fibers for efficiently collecting reflectance, fluorescence, and Raman spectra from the same tissue location. A probe with this excitation/collection geometry has not been used previously to collect reflectance and fluorescence spectra, and thus physical tissue models (“phantoms”) are used to characterize the probe’s spectroscopic response. This calibration provides probe-specific modeling parameters that enable accurate extraction of spectral parameters. This clinical MMS system has been used recently to analyze artery and breast tissue in vivo and ex vivo.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Spectroscopy Laboratory
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1063/1.3117832