Repository logo
Log in(current)
Repository logoMIT Open ScholarshipDSpace@MIT
  1. Home
  2. MIT Open Access Articles
  3. MIT Open Access Articles
  4. Conceptual Study for Tissue-Regenerative Biodegradable Magnesium Implant Integrated with Nitric Oxide-Releasing Nanofibers

Conceptual Study for Tissue-Regenerative Biodegradable Magnesium Implant Integrated with Nitric Oxide-Releasing Nanofibers

Thumbnail Image
Download
Name

12540_2018_232_ReferencePDF.pdf

Size

842.04 KB

Format

Adobe PDF

Checksum (MD5)

6f6532fbfc6860bda0719c9edbe1cad7

Download all files submitted through automated deposit
art_7815390282158639110.zip (729.69 KB)
Author(s)
Jeon, Jin-Kyung
•
Seo, Hyunseon
•
Park, Jimin
•
Son, Soo Ji
•
Kim, Yeong Rim
•
Kim, Eun Shil
•
Park, Jong Woong
•
Jung, Woong-Gyo
•
Jeon, Hojeong
•
Kim, Yu-Chan
more
Date Issued
January 2019
Journal
Metals and Materials International
Publisher
Springer Science and Business Media LLC
Citation
Jeon, Jin-Kyung et al. "Conceptual Study for Tissue-Regenerative Biodegradable Magnesium Implant Integrated with Nitric Oxide-Releasing Nanofibers." Metals and Materials International 25, 4 (January 2018): 1098–1107 © 2019 The Korean Institute of Metals and Materials
Abstract
The excessive initial corrosion rate of Mg is a critical limitation in the clinical application of biodegradable Mg implants because the device loses its fixation strength before the fractured bone heals. This study suggests a new approach to overcome this hurdle by accelerating tissue regeneration instead of delaying the implant biodegradation. As angiogenesis is an essential process in early bone regeneration, a Mg implant coated with electrospun nanofibers containing nitric oxide (NO), which physiologically promotes angiogenesis, is designed. The integrated device enables adjustable amounts of NO to be stored on the NO donor-conjugated nanofiber coating, stably delivered, and released to the fractured bone tissue near the implanted sites. An in vitro corrosion test reveals no adverse effect of the released NO on the corrosion behavior of the Mg implant. Simultaneously, the optimal concentration level of NO released from the implant significantly enhances tube network formation of human umbilical vein endothelial cells without any cytotoxicity problem. This indicates that angiogenesis can be accelerated by combining NO-releasing nanofibers with a Mg implant. With its proven feasibility, the proposed approach could be a novel solution for the initial stability problem of biodegradable Mg implants, leading to successful bone fixation.
MIT Department
Massachusetts Institute of Technology. Department of Materials Science and Engineering
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.