Automated Quantitative Analysis of Capnogram Shape for COPD–Normal and COPD–CHF Classification
Name
IEEE Xplore Full-Text PDF_.pdf
Size
2.45 MB
Format
Adobe PDF
Checksum (MD5)
ccb7dbcabc66ff976b86d02eaa08bb87
Author(s) • • • • • • •
Mieloszyk, Rebecca J.
Verghese, George, C.
Deitch, Kenneth
Cooney, Brendan
Khalid, Abdullah
Mirre-Gonzalez, Milciades A.
Heldt, Thomas
Krauss, Baruch S.
Date Issued
June 2014
Journal
IEEE Transactions on Biomedical Engineering
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Mieloszyk, Rebecca J., George C. Verghese, Kenneth Deitch, Brendan Cooney, Abdullah Khalid, Milciades A. Mirre-Gonzalez, Thomas Heldt, and Baruch S. Krauss. "Automated Quantitative Analysis of Capnogram Shape for COPD–Normal and COPD–CHF Classification." IEEE Transactions on Biomedical Engineering, (Volume:PP, Issue:99). p.1-9.
Version
Author's final manuscript
Abstract
We develop an approach to quantitative analysis of carbon dioxide concentration in exhaled breath, recorded as a function of time by capnography. The generated waveform – or capnogram – is currently used in clinical practice to establish the presence of respiration as well as determine respiratory rate and end-tidal CO2 concentration. The capnogram shape also has diagnostic value, but is presently assessed qualitatively, by visual inspection. Prior approaches to quantitatively characterizing the capnogram shape have explored the correlation of various geometric parameters with pulmonary function tests. These studies attempted to characterize the capnogram in normal subjects and patients with cardiopulmonary disease, but no consistent progress was made, and no translation into clinical practice was achieved. We apply automated quantitative analysis to discriminate between chronic obstructive pulmonary disease (COPD) and congestive heart failure (CHF), and between COPD and normal. Capnograms were collected from 30 normal subjects, 56 COPD patients, and 53 CHF patients. We computationally extract four physiologically based capnogram features. Classification on a hold-out test set was performed by an ensemble of classifiers employing quadratic discriminant analysis, designed through cross-validation on a labeled training set. Using 80 exhalations of each capnogram record in the test set, performance analysis with bootstrapping yields areas under the receiver operating characteristic (ROC) curve of 0.89 (95% CI: 0.72-0.96) for COPD/CHF classification, and 0.98 (95% CI: 0.82- 1.0) for COPD/normal classification. This classification performance is obtained with a run-time sufficiently fast for realtime monitoring.
MIT Department
Institute for Medical Engineering and Science
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Research Laboratory of Electronics
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1109/TBME.2014.2332954