Comparing the quantification of Forster resonance energy transfer measurement accuracies based on intensity, spectral, and lifetime imaging
Name
Pelet-2006-Comparing the quantification.pdf
Size
730.84 KB
Format
Adobe PDF
Checksum (MD5)
71ffa222f0da257ae40ed80c5a73fbae
Author(s) • •
Pelet, Serge
Previte, Michael J. R.
So, Peter T. C.
Date Issued
May 2006
Journal
Journal of Biomedical Optics
Publisher
SPIE
Citation
Pelet, Serge, Michael J. R. Previte, and Peter T. C. So. “Comparing the Quantification of Forster Resonance Energy Transfer Measurement Accuracies Based on Intensity, Spectral, and Lifetime Imaging.” Journal of Biomedical Optics 11, no. 3 (2006): 034017. © 2006 Society of Photo-Optical Instrumentation Engineers
Version
Final published version
Abstract
The measurement of Förster resonance energy transfer (FRET) in microscopes can be realized by different imaging modalities. In the present work, reference FRET constructs are developed to allow the comparison of FRET microscopy measurements using intensity, spectral, and lifetime imaging. Complimentary DNA strands are respectively labeled with Oregon Green 488 (OG488) or tetramethylrhodamine (TMR). The OG488 dye is fixed at the 5′ end of one strand, and the TMR label position is allowed to vary along the complimentary strand. Since OG488 and TMR are FRET pairs, the FRET efficiency can be determined theoretically from the distance separating the two dyes of the double-stranded DNA molecules. Microscopic images are formed by imaging microcapillaries containing various mixtures of oligonucleotides labeled with the FRET fluorophore pair, only the donor, or only acceptor. Traditional two-channel intensity measurements are compared with spectrally resolved imaging and fluorescence lifetime imaging by calculating a FRET index. The latter proves to be the best method to quantify FRET efficiency in the image. More importantly, the intensity fraction of molecules undergoing FRET can be quantitatively measured in each pixel of the image.
MIT Department
Massachusetts Institute of Technology. Department of Biological Engineering
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1117/1.2203664