Alignment of the CMS muon system with cosmic-ray and beam-halo muons
Name
0911.4022.pdf
Description
Accepted version
Size
1.34 MB
Format
Adobe PDF
Checksum (MD5)
5567eeea1a2b10a7a32296977d9b4a80
Author(s) • • • • • • • • •
Alver, Burak Han
Bauer, Gerry P
Bendavid, Joshua L.
Busza, Wit
Butz, Erik M.
Cali, Ivan Amos
Chan, M.
D'Enterria, David
Everaerts, Pieter Bruno Bart
Gomez-Ceballos, Guillelmo
Date Issued
March 2010
Journal
Journal of Instrumentation
Publisher
IOP Publishing
Citation
CMS Collaboration et al. "Alignment of the CMS muon system with cosmic-ray and beam-halo muons." Journal of Instrumentation, 5 (March 2010): T03020 © 2010 IOP Publishing Ltd and SISSA
Version
Author's final manuscript
Abstract
The CMS muon system has been aligned using cosmic-ray muons collected in 2008 and beam-halo muons from the 2008 LHC circulating beam tests. After alignment, the resolution of the most sensitive coordinate is 80 microns for the relative positions of superlayers in the same barrel chamber and 270 microns for the relative positions of endcap chambers in the same ring structure. The resolution on the position of the central barrel chambers relative to the tracker is comprised between two extreme estimates, 200 and 700 microns, provided by two complementary studies. With minor modifications, the alignment procedures can be applied using muons from LHC collisions, leading to additional significant improvements.
MIT Department
Massachusetts Institute of Technology. Department of Physics
Massachusetts Institute of Technology. Department of Nuclear Science and Engineering
Massachusetts Institute of Technology. Laboratory for Nuclear Science
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1088/1748-0221/5/03/T03020