LOW-COST BACTERIAL DETECTION SYSTEM FOR FOOD SAFETY BASED ON AUTOMATED DNA EXTRACTION, AMPLIFICATION AND READOUT
Name
uTAS-13_FoodSafety_0271.pdf
Size
935.47 KB
Format
Adobe PDF
Checksum (MD5)
5f03cf9f88bbfdd9c15e0cd375956e21
Author(s) • • • • •
Hoehl, Melanie Margarete
Bocholt, Eva Schulte
Karippai, Nobu
Zengerle, Roland
Steigert, Juergen
Slocum, Alexander H.
Date Issued
October 2013
Journal
Proceedings of the 17th International Conference on Miniaturized Systems for chemistry and Life Sciences (MicroTAS 2013)
Publisher
Chemical and Biological Microsystems Society
Citation
Hoehl, Melanie, Eva Schulte Bocholt, Nobu Karippai, Roland Zengerle, Juergen Steigert, and Alexander Slocum. "LOW-COST BACTERIAL DETECTION SYSTEM FOR FOOD SAFETY BASED ON AUTOMATED DNA EXTRACTION, AMPLIFICATION AND READOUT." In 17th International Conference on Miniaturized Systems for Chemistry and Life Sciences (µTAS 2013), Messe Freiburg in Freiburg, GERMANY, 27-31 October 2013.
Version
Author's final manuscript
Abstract
To ensure food, medical and environmental safety and quality, rapid, low-cost and easy-to-use detection methods are desirable. Here, the LabSystem is introduced for integrated, automated DNA purification and amplification. It consists of a disposable, centrifugally-driven DNA purification platform (LabTube) and the subsequent amplification in a low-cost UV/vis-reader (LabReader). In this paper, food safety was chosen as the first sample application with pathogenic verotoxin-producing (VTEC) Escherichia coli (E.coli) in water and milk, and the product-spoiler Alicyclobacillus acidoterrestris (A acidoterrestris) in apple juice as sample organisms. DNA was amplified qualitatively using isothermal loop-mediated DNA amplification (LAMP) and quantitatively using real-time PCR. By optimizing manual purification protocols inside the LabTube, as little as 45 inserted DNA copies were extracted from E.coli and A.acidoterrestris lysates in real samples (milk, juice and water). To run isothermal DNA amplification (LAMP) and PCR inside the LabReader, temperature control as well as data analysis methods were implemented. Combined detection limits for DNA purification and amplification from bacteria lysates in real samples at 102-103 inserted copies were achieved. The demonstrated LabSystem runs with standard laboratory equipment and reduces hands-on times from 30min to 3min.
MIT Department
Harvard University--MIT Division of Health Sciences and Technology
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike 3.0
Persistent DSpace Link
DOI of Published Version
http://www.microtas2013.org/program/MicroTAS2013_TechnicalProgram.pdf