Acoustic Method for Detection Large Pneumothoraxes in the Out-of-Hospital Environment
Name
motz-amotz-sb-meche-2023-thesis.pdf
Description
Thesis PDF
Size
1.7 MB
Format
Adobe PDF
Checksum (MD5)
d7e6f926f7311b40c4b03b86ddb2c6f6
Author(s)
Motz, Andrew J.
Advisor(s)
Hanumara, Nevan
Date Issued
June 2023
Publisher
Massachusetts Institute of Technology
Abstract
Pneumothoraxes and hemothoraxes present a medical emergency that can easily be life threatening. However, improper and unnecessarily aggressive treatment is not uncommon and poses significant risk to patients. Much of this is a result of poor diagnostic aids to assist in the detection of pneumothoraxes out-of-hospital, such as by Emergency Medical Service (EMS) providers. A simple, portable, and accurate method of clinically detecting a pneumothorax would increase the quality of patient care in austere environments. We propose to evaluate the acoustic transfer function of the human thorax for signs that a pocket of air or blood has accumulated within the pleural space. In an attempt to validate this hypothesis, we constructed a phantom model of the human thorax with artificial lungs and pleural spaces capable of being inflated with variable sized air pockets to displace the lungs. The model was subjected to a known acoustic input in the form of a frequency sweep and a low frequency square wave and the output signal was recorded at multiple points on the phantom. Analysis of four characteristics of the output revealed no discernable correlation to the volume or location of a pneumothorax. While the initial attempt to verify the hypothesis was unsuccessful, the theory remains intact. Future work to further this hypothesis could include improving the mathematical model and/or creating a higher fidelity phantom model for experimentation.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
In Copyright - Educational Use Permitted
Copyright retained by author(s)
Persistent DSpace Link