Realization of a 15-channel, hermetically-encased wireless subretinal for the blind
Name
Kelly-2009-Realization of a 15-.pdf
Size
853.16 KB
Format
Adobe PDF
Checksum (MD5)
039c0a69ffa1a77f91292781345db42d
Author(s) • • • • • • • • •
Wyatt, John L.
Drohan, William A.
Doyle, Patrick S.
Kelly, Shawn K.
Rizzo, Joseph F., III
Cogan, Stuart F.
Theogarajan, Luke S.
Gingerich, Marcus D.
Chen, Jinghua
Shire, Douglas B.
Date Issued
November 2009
Journal
2009 Engineering in Medicine and Biology Society
Publisher
Institute of Electrical and Electronics Engineers
Citation
Kelly, S.K. et al. “Realization of a 15-channel, hermetically-encased wireless subretinal prosthesis for the blind.” Engineering in Medicine and Biology Society, 2009. EMBC 2009. Annual International Conference of the IEEE. 2009. 200-203. © 2009 IEEE
Version
Final published version
Abstract
A miniaturized, hermetically-encased, wirelessly-operated retinal prosthesis has been developed for implantation and pre-clinical studies in Yucatan mini-pig animal models. The prosthesis conforms to the eye and drives a microfabricated polyimide stimulating electrode array with sputtered iridium oxide electrodes. This array is implanted in the subretinal space using a specially-designed ab externo surgical technique that affixes the bulk of the prosthesis to the surface of the sclera. The implanted device includes a hermetic titanium case containing a 15-channel stimulator chip and discrete power supply components. Feedthroughs from the case connect to secondary power- and data-receiving coils. In addition, long-term in vitro pulse testing was performed on the electrodes to ensure their stability for the long lifetime of the hermetic case. The final assembly was tested in vitro to verify wireless operation of the system in biological saline using a custom RF transmitter circuit and primary coils. Stimulation pulse strength, duration and frequency were programmed wirelessly using a custom graphical user interface. Operation of the retinal implant has been verified in vivo in one pig for more than three months by measuring stimulus artifacts on the eye surface using a contact lens electrode.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Research Laboratory of Electronics
Terms of Use
Article is made available in accordance with the publisher’s policy and may be subject to US copyright law. Please refer to the publisher’s site for terms of use.
Persistent DSpace Link
DOI of Published Version
http://dx.doi.org/10.1109/IEMBS.2009.5333619