A Magnetic Retrieval System for Stents in the Pancreaticobiliary Tree
Name
Cantillon-Murphy-2010-A Magnetic Retrieval.pdf
Size
319.56 KB
Format
Adobe PDF
Checksum (MD5)
16e6d65e1bee0706d2f579582dc7f834
Author(s) • • • • • •
Cantillon-Murphy, Padraig
Ryou, Marvin
Shaikh, Sohail N.
Azagury, Dan
Ryan, Michele
Thompson, Christopher C.
Lang, Jeffrey H.
Date Issued
August 2010
Journal
IEEE Transactions on Biomedical Engineering
Publisher
Institute of Electrical and Electronics Engineers
Citation
Cantillon-Murphy, P. et al. “A Magnetic Retrieval System for Stents in the Pancreaticobiliary Tree.” Biomedical Engineering, IEEE Transactions on 57.8 (2010): 2018-2025. © 2011 IEEE.
Version
Final published version
Abstract
Clinical endoscopic intervention of the pancreaticobiliary tree [endoscopic retrograde cholangiopancreatography (ERCP)] often concludes with the insertion of a temporary plastic stent to reduce the risk of post-ERCP complications by promoting continued flow of bile and pancreatic fluids. This stent is later removed once the patient has fully recovered, but today this necessitates a second endoscopic intervention. The final goal of this work is to obviate the second intervention. This is to be achieved by adding a magnetic ring to the stent such that the stent is removed using a hand-held magnet, held in a suitable position ex vivo . This paper details the design, optimization, and both ex vivo and in vivo testing of the magnetized stent and hand-held magnet, which has been accomplished to date. The optimized design for the hand-held magnet and the modified stent with a magnetic attachment performs in line with simulated expectations, and successful retrieval is achieved in the porcine ex vivo setting at 9-10 cm separation. This is comparable to the mean target capture distance of 10 cm between the entry point to the biliary system and the closest cutaneous surface, determined from random review of clinical fluoroscopies in ten human patients. Subsequently, the system was successfully tested in vivo in the acute porcine model, where retrieval at an estimated separation of 5-6 cm was captured on endoscopic video. These initial results indicate that the system may represent a promising approach for the elimination of a second endoscopic procedures following placement of pancreatic and biliary stents.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1109/tbme.2010.2045653