  4. In-situ calibration of the water vapor channel for multi-filter rotating shadowband radiometer using collocated GPS, AERONET and meteorology data

In-situ calibration of the water vapor channel for multi-filter rotating shadowband radiometer using collocated GPS, AERONET and meteorology data

Thumbnail Image
Download
Name

99750E.pdf

Size

757.71 KB

Format

Adobe PDF

Checksum (MD5)

e69e0f44b02e944cddf36792624c3766

Download all files submitted through automated deposit
sword-2018-03-16.original.zip (713.13 KB)
Author(s)
Chen, Maosi
•
Zempila, Melina-Maria
•
Davis, John M.
•
Gao, Wei
•
King, Robert W
Date Issued
August 2016
Journal
Proceedings Volume 9975, Remote Sensing and Modeling of Ecosystems for Sustainability XIII
Publisher
SPIE
Citation
Chen, Maosi, et al. "In-Situ Calibration of the Water Vapor Channel for Multi-Filter Rotating Shadowband Radiometer Using Collocated GPS, AERONET and Meteorology Data." Proceedings Volume 9975, Remote Sensing and Modeling of Ecosystems for Sustainability XIII, 28 August 1, - September, 2016, San Diego, California, edited by Wei Gao and Ni-Bin Chang, SPIE, 2016, p. © COPYRIGHT SPIE.
Version
Final published version
Abstract
The difficulty of in-situ calibration on the 940 nm channel of Multi-Filter Rotating Shadowband Radiometer (MFRSR) stems from the distinctive non-linear relationship between the amount of precipitable water vapor (PW) and its optical depth (i.e. curve of growth) compared to the counterpart of aerosols. Previous approaches, the modified Langley methods (MLM), require exact aerosol optical depth (AOD) values and a constant PW value at all points participating the regression. Instead, we propose a new method that substitutes the PW optical depth derived from collocated GPS zenith wet delay retrieval in conjunction with meteorology data and requires a constant AOD value at all points participating the regression. The main benefits of the new method include: (1) Aerosol stability is easier to fulfill than PW stability; (2) AOD stability could be inferred from adjacent channels (e.g. 672 and 870 nm) of MFRSR itself without measurements of a collocated AERONET sun photometer; and (3) When applicable, the time interval of GPS derived PW (i.e. 3 minutes) is more compatible with the MFRSR sampling interval (i.e. 3 minutes) than AERONET interpolated AOD (i.e. 15 minutes). Both MLM and the new method were applied to the MFRSR of USDA UV-B Monitoring and Research Program at the station in Billings, Oklahoma (active for 18 years so far) on July 28, 2015. The performances of the two methods are compared in order to assess their accuracy and the advantages and disadvantages.
MIT Department
Massachusetts Institute of Technology. Department of Earth, Atmospheric, and Planetary Sciences
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
http://hdl.handle.net/1721.1/116507
DOI of Published Version
https://doi.org/10.1117/12.2236572