  4. Fast reconstruction for multichannel compressed sensing using a hierarchically semiseparable solver

Fast reconstruction for multichannel compressed sensing using a hierarchically semiseparable solver

Thumbnail Image
Download
Name

Fast reconstruction.pdf

Size

773.76 KB

Format

Adobe PDF

Checksum (MD5)

a7f15564469fd3b18c577cb50639befb

Author(s)
Cauley, Stephen F.
•
Xi, Yuanzhe
•
Bilgic, Berkin
•
Xia, Jianlin
•
Balakrishnan, Venkataramanan
•
Setsompop, Kawin
•
Adalsteinsson, Elfar
•
Wald, Lawrence
Date Issued
February 2015
Journal
Magnetic Resonance in Medicine
Publisher
Wiley Blackwell
Citation
Cauley, Stephen F.; Xi, Yuanzhe; Bilgic, Berkin et al. “Fast Reconstruction for Multichannel Compressed Sensing Using a Hierarchically Semiseparable Solver.” Magnetic Resonance in Medicine 73, 3 (March 2014): 1034–1040 © 2014 Wiley Periodicals, Inc
Version
Author's final manuscript
Abstract
Purpose The adoption of multichannel compressed sensing (CS) for clinical magnetic resonance imaging (MRI) hinges on the ability to accurately reconstruct images from an undersampled dataset in a reasonable time frame. When CS is combined with SENSE parallel imaging, reconstruction can be computationally intensive. As an alternative to iterative methods that repetitively evaluate a forward CS+SENSE model, we introduce a technique for the fast computation of a compact inverse model solution. Methods A recently proposed hierarchically semiseparable (HSS) solver is used to compactly represent the inverse of the CS+SENSE encoding matrix to a high level of accuracy. To investigate the computational efficiency of the proposed HSS-Inverse method, we compare reconstruction time with the current state-of-the-art. In vivo 3T brain data at multiple image contrasts, resolutions, acceleration factors, and number of receive channels were used for this comparison. Results The HSS-Inverse method allows for math formula speedup when compared to current state-of-the-art reconstruction methods with the same accuracy. Efficient computational scaling is demonstrated for CS+SENSE with respect to image size. The HSS-Inverse method is also shown to have minimal dependency on the number of parallel imaging channels/acceleration factor. Conclusions The proposed HSS-Inverse method is highly efficient and should enable real-time CS reconstruction on standard MRI vendors' computational hardware.
MIT Department
Harvard University--MIT Division of Health Sciences and Technology
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
http://creativecommons.org/licenses/by-nc-sa/4.0/
Persistent DSpace Link
http://hdl.handle.net/1721.1/110733
DOI of Published Version
https://doi.org/10.1002/mrm.25222
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.