Impact of flow pulsatility on arterial drug distribution in stent-based therapy
Name
Edelman_Impact of flow.pdf
Size
3.48 MB
Format
Adobe PDF
Checksum (MD5)
faa29b66b9c53ebcc33ba9ba2a6261cd
Author(s) • • • •
O’Brien, Caroline C.
Kolachalama, Vijaya Bhasker
Barber, Tracie J.
Simmons, Anne
Edelman, Elazer R.
Date Issued
June 2013
Journal
Journal of Controlled Release
Publisher
Elsevier B.V.
Citation
O’Brien, Caroline C., Vijaya B. Kolachalama, Tracie J. Barber, Anne Simmons, and Elazer R. Edelman. “Impact of Flow Pulsatility on Arterial Drug Distribution in Stent-Based Therapy.” Journal of Controlled Release 168, no. 2 (June 2013): 115–124.
Version
Author's final manuscript
Abstract
Drug-eluting stents reside in a dynamic fluid environment where the extent to which drugs are distributed within the arterial wall is critically modulated by the blood flowing through the arterial lumen. Yet several factors associated with the pulsatile nature of blood flow and their impact on arterial drug deposition have not been fully investigated. We employed an integrated framework comprising bench-top and computational models to explore the factors governing the time-varying fluid dynamic environment within the vasculature and their effects on arterial drug distribution patterns. A custom-designed bench-top framework comprising a model of a single drug-eluting stent strut and a poly-vinyl alcohol-based hydrogel as a model tissue bed simulated fluid flow and drug transport under fully apposed strut settings. Bench-top experiments revealed a relative independence between drug distribution and the factors governing pulsatile flow and these findings were validated with the in silico model. Interestingly, computational models simulating suboptimal deployment settings revealed a complex interplay between arterial drug distribution, Womersley number and the extent of malapposition. In particular, for a stent strut offset from the wall, total drug deposition was sensitive to changes in the pulsatile flow environment, with this dependence increasing with greater wall displacement. Our results indicate that factors governing pulsatile luminal flow on arterial drug deposition should be carefully considered in conjunction with device deployment settings for better utilization of drug-eluting stent therapy.
MIT Department
Institute for Medical Engineering and Science
Harvard University--MIT Division of Health Sciences and Technology
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1016/j.jconrel.2013.03.014