  4. Quantification of thrombus formation in malapposed coronary stents deployed in vitro through imaging analysis

Quantification of thrombus formation in malapposed coronary stents deployed in vitro through imaging analysis

Thumbnail Image
Download
Name

nihms942879.pdf

Description
Accepted version
Size

581.79 KB

Format

Adobe PDF

Checksum (MD5)

5e2c36d56b8ef8c6d2c7193c8baeb600

sword-2019-10-09T18:09:41.original.xml (130 B)
Original SWORD entry document
Author(s)
Brown, Jonathan
•
O'Brien, Caroline C.
•
Lopes Jr, Augusto Celso de Araujo
•
Kolandaivelu, Kumaran
•
Edelman, Elazer R
Date Issued
April 2018
Journal
Journal of Biomechanics
Publisher
Elsevier BV
Citation
Brown, Jonathan et al. "Quantification of thrombus formation in malapposed coronary stents deployed in vitro through imaging analysis." Journal of Biomechanics 71 (April 2018): 296-301 © 2018 Elsevier Ltd
Version
Author's final manuscript
Abstract
Stent thrombosis is a major complication of coronary stent and scaffold intervention. While often unanticipated and lethal, its incidence is low making mechanistic examination difficult through clinical investigation alone. Thus, throughout the technological advancement of these devices, experimental models have been indispensable in furthering our understanding of device safety and efficacy. As we refine model systems to gain deeper insight into adverse events, it is equally important that we continue to refine our measurement methods. We used digital signal processing in an established flow loop model to investigate local flow effects due to geometric stent features and ultimately its relationship to thrombus formation. A new metric of clot distribution on each microCT slice termed normalized clot ratio was defined to quantify this distribution. Three under expanded coronary bare-metal stents were run in a flow loop model to induce clotting. Samples were then scanned in a MicroCT machine and digital signal processing methods applied to analyze geometric stent conformation and spatial clot formation. Results indicated that geometric stent features play a significant role in clotting patterns, specifically at a frequency of 0.6225 Hz corresponding to a geometric distance of 1.606 mm. The magnitude-squared coherence between geometric features and clot distribution was greater than 0.4 in all samples. In stents with poor wall apposition, ranging from 0.27 mm to 0.64 mm maximum malapposition (model of real-world heterogeneity), clots were found to have formed in between stent struts rather than directly adjacent to struts. This early work shows how the combination of tools in the areas of image processing and signal analysis can advance the resolution at which we are able to define thrombotic mechanisms in in vitro models, and ultimately, gain further insight into clinical performance.
MIT Department
Massachusetts Institute of Technology. Institute for Medical Engineering & Science
Terms of Use
Creative Commons Attribution-NonCommercial-NoDerivs License
http://creativecommons.org/licenses/by-nc-nd/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/126400
DOI of Published Version
https://doi.org/10.1016/j.jbiomech.2018.01.044
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.