  4. Accuracy of cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm to identify sites of origin of cardiac electrical activation

Accuracy of cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm to identify sites of origin of cardiac electrical activation

Thumbnail Image
Download
Name

10840_2019_605_ReferencePDF.pdf

Size

1.5 MB

Format

Adobe PDF

Checksum (MD5)

36001e9d8179b38122354c78de03ff0c

Download all files submitted through automated deposit
art_5011415450902278856.zip (1.31 MB)
Author(s)
Lv, Wener
•
Lee, Kichang
•
Arai, Tatsuya
•
Barrett, Conor D
•
Hasan, Maysun Mazhar
•
Hayward, Alison M
•
Marini, Robert P.
•
Barley, Maya E.
•
Galea, Anna
•
Hirschman, Gordon
more
Date Issued
September 2020
Journal
Journal of Interventional Cardiac Electrophysiology
Publisher
Springer US
Citation
Lv, Wener et al. "Accuracy of cardiac ablation catheter guidance by means of a single equivalent moving dipole inverse algorithm to identify sites of origin of cardiac electrical activation." Journal of Interventional Cardiac Electrophysiology 58 (September 2020): 323-31 ©2020 Author(s)
Version
Author's final manuscript
Abstract
We have developed a system that could potentially be used to identify the site of origin of ventricular tachycardia (VT) and to guide a catheter to that site to deliver radio-frequency ablation therapy. This system employs the Inverse Solution Guidance Algorithm based upon Single Equivalent Moving Dipole (SEMD) localization method. The system was evaluated in in vivo swine experiments. Arrays consisting of 9 or 16 bipolar epicardial electrodes and an additional mid-myocardial pacing lead were sutured to each ventricle. Focal tachycardia was simulated by applying pacing pulses to each epicardial electrode at multiple pacing rates during breath hold at the end-expiration phase. Surface potentials were recorded from 64 surface electrodes and then analyzed using the SEMD method to localize the position of the pacing electrodes. We found a close correlation between the locations of the pacing electrodes as measured in computational and real spaces. The reproducibility error of the SEMD estimation of electrode location was 0.21 ± 0.07 cm. The vectors between every pair of bipolar electrodes were computed in computational and real spaces. At 120 bpm, the lengths of the vectors in the computational and real space had a 95% correlation. Computational space vectors were used in catheter guidance simulations which showed that this method could reduce the distance between the real space locations of the emulated catheter tip and the emulated arrhythmia origin site by approximately 72% with each movement. We have demonstrated the feasibility of using our system to guide a catheter to the site of the emulated VT origin.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Massachusetts Institute of Technology. Institute for Medical Engineering & Science
Massachusetts Institute of Technology. Department of Aeronautics and Astronautics
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Division of Comparative Medicine
Harvard University--MIT Division of Health Sciences and Technology
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
http://creativecommons.org/licenses/by-nc-sa/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/127822
DOI of Published Version
https://doi.org/10.1007/s10840-019-00605-z