Repository logo
Log in(current)
Repository logoMIT Open ScholarshipDSpace@MIT
  1. Home
  2. MIT Open Access Articles
  3. MIT Open Access Articles
  4. Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs

Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs

Thumbnail Image
Download
Name

023502_1_5.0206206.pdf

Description
Published version
Size

5.94 MB

Format

Adobe PDF

Checksum (MD5)

d897323361776d211eacef3a5dfac4fc

sword-2026-04-07T15:43:53.original.xml (130 B)
Original SWORD entry document
Author(s)
Johnson, TM
•
Lahmann, B
•
Russell, L
•
Vanderloo, NL
•
Cufari, MJ
•
Reichelt, BL
•
Chang, CW
•
Birkel, A
•
Kabadi, NV
•
Sutcliffe, GD
more
Date Issued
February 4, 2025
Journal
Review of Scientific Instruments
Publisher
AIP Publishing
Citation
T. M. Johnson, B. Lahmann, L. Russell, N. L. Vanderloo, M. J. Cufari, B. L. Reichelt, C. W. Chang, A. Birkel, N. V. Kabadi, G. D. Sutcliffe, P. J. Adrian, J. A. Pearcy, J. H. Kunimune, S. G. Dannhoff, T. E. Evans, M. Gatu Johnson, F. H. Séguin, R. D. Petrasso, C. K. Li, J. A. Frenje; Monte Carlo toolkit for designing and validating step-range-filter spectrometer designs. Rev. Sci. Instrum. 1 February 2025; 96 (2): 023502.
Version
Final published version
Abstract
Here, we present a Monte Carlo toolkit for validating step range filter (SRF) spectrometer designs. Geant4 is used to transport charged particles through the SRF filters to generate synthetic SRF data that include realistic CR-39 effects. Synthetic SRF spectra generated by this method inherently account for instrument response and allow for the quantification of SRF performance before shots. The usefulness of this toolkit is demonstrated through its application to a number of problems. A new broadband SRF for the ∼10 MeV wide 3He3He proton spectrum is validated, and an analysis method for analyzing 3He3He-p SRF data that accounts for instrument response is put forth. In addition, an SRF design for the compact recoil-proton spectrometer (CRS) on the Z-machine is validated. Finally, a new calibration technique for the DD-p SRF is proposed and validated.
MIT Department
Massachusetts Institute of Technology. Plasma Science and Fusion Center
Terms of Use
Creative Commons Attribution
https://creativecommons.org/licenses/by/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/165355
DOI of Published Version
https://doi.org/10.1063/5.0206206
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.