Human dopamine receptor nanovesicles for gate-potential modulators in high-performance field-effect transistor biosensors
Name
Park-2014-Human dopamine receptor.pdf
Size
1.03 MB
Format
Adobe PDF
Checksum (MD5)
c1dff1c70d5b08ad71f03536703745d4
Author(s) • • • • • • • • •
Park, Seon Joo
Song, Hyun Seok
Kwon, Oh Seok
Chung, Ji Hyun
Lee, Seung Hwan
An, Ji Hyun
Ahn, Sae Ryun
Lee, Ji Eun
Yoon, Hyeonseok
Park, Tai Hyun
Date Issued
March 2014
Journal
Scientific Reports
Publisher
Nature Publishing Group
Citation
Park, Seon Joo, Hyun Seok Song, Oh Seok Kwon, Ji Hyun Chung, Seung Hwan Lee, Ji Hyun An, Sae Ryun Ahn, et al. “Human Dopamine Receptor Nanovesicles for Gate-Potential Modulators in High-Performance Field-Effect Transistor Biosensors.” Sci. Rep. 4 (March 11, 2014).
Version
Final published version
Abstract
The development of molecular detection that allows rapid responses with high sensitivity and selectivity remains challenging. Herein, we demonstrate the strategy of novel bio-nanotechnology to successfully fabricate high-performance dopamine (DA) biosensor using DA Receptor-containing uniform-particle-shaped Nanovesicles-immobilized Carboxylated poly(3,4-ethylenedioxythiophene) (CPEDOT) NTs (DRNCNs). DA molecules are commonly associated with serious diseases, such as Parkinson's and Alzheimer's diseases. For the first time, nanovesicles containing a human DA receptor D1 (hDRD1) were successfully constructed from HEK-293 cells, stably expressing hDRD1. The nanovesicles containing hDRD1 as gate-potential modulator on the conducting polymer (CP) nanomaterial transistors provided high-performance responses to DA molecule owing to their uniform, monodispersive morphologies and outstanding discrimination ability. Specifically, the DRNCNs were integrated into a liquid-ion gated field-effect transistor (FET) system via immobilization and attachment processes, leading to high sensitivity and excellent selectivity toward DA in liquid state. Unprecedentedly, the minimum detectable level (MDL) from the field-induced DA responses was as low as 10 pM in real- time, which is 10 times more sensitive than that of previously reported CP based-DA biosensors. Moreover, the FET-type DRNCN biosensor had a rapid response time (<1 s) and showed excellent selectivity in human serum.
MIT Department
Institute for Medical Engineering and Science
Harvard University--MIT Division of Health Sciences and Technology
Massachusetts Institute of Technology. Department of Chemistry
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1038/srep04342