Motion and Sash Height (MASH) alarms for efficient fume hood use
Name
s41598-021-00772-y.pdf
Description
Accepted version
Size
2.43 MB
Format
Unknown
Checksum (MD5)
5e1e3c75731f3f256fd6a5dcf4fde562
Author(s) • • •
Kongoletos, Johnathan
Munden, Ethan
Ballew, Jennifer
Preston, Daniel John
Date Issued
December 2021
Journal
Scientific Reports
Publisher
Springer Science and Business Media LLC
Citation
Kongoletos, Johnathan, Munden, Ethan, Ballew, Jennifer and Preston, Daniel J. 2021. "Motion and Sash Height (MASH) alarms for efficient fume hood use." Scientific Reports, 11 (1).
Version
Final published version
Abstract
Ventilation, including fume hoods, consumes 40–70% of the total energy used by modern laboratories. Energy-conscious fume hood usage—for example, closing the sash when a hood is unused—can significantly reduce energy expenditures due to ventilation. Prior approaches to promote such behaviors among lab users have primarily relied on passive feedback methods. In this work, we developed a low-cost fume hood monitoring device with active feedback to alert lab users when a fume hood is left open and unused. Using data collected by the building management system, we observed a 75.6% decrease in the average sash height after installation of these “Motion and Sash Height” (MASH) alarms, which would result in a reduction roughly equal to 43% of the annual carbon emissions of a typical American vehicle, for each fume hood. The MASH alarm presented here reduced energy costs by approximately $1,159 per year, per hood, at MIT.
MIT Department
Massachusetts Institute of Technology. Department of Architecture
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
Creative Commons Attribution 4.0 International license
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1038/s41598-021-00772-y