An ultrasound probe sheathing device : design, development, and evaluation
Name
858868528-MIT.pdf
Description
Full printable version
Size
12.2 MB
Format
Adobe PDF
Checksum (MD5)
66979ac0d9b48fed5cdf3b91be1c454a
Author(s)
Hernley, Lauren R. (Lauren Renee)
Advisor(s)
David R. Wallace.
Date Issued
2013
Publisher
Massachusetts Institute of Technology
Abstract
During interventional ultrasound-guided procedures, sterility is maintained by covering the ultrasound transducer head and cord with a sterile sheath. The current sheathing technique is cumbersome, requires an assistant to complete, and poses a risk of tangling the probe cord and breaching the sterile barrier. This paper presents the design, development, and evaluation of a probe holder and cartridge-style, single-use applicator that returns high quality ultrasound images and sheaths ultrasound transducers 35% faster than the current technique with a decreased risk of breaking sterility. When tested by medical professionals, all participants preferred the new device to the existing probe sheathing method.
Description
Thesis (S.M.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2013.
Cataloged from PDF version of thesis.
Includes bibliographical references (p. 91-92).
Subjects
Mechanical Engineering.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
M.I.T. theses are protected by
copyright. They may be viewed from this source for any purpose, but
reproduction or distribution in any format is prohibited without written
permission. See provided URL for inquiries about permission.
copyright. They may be viewed from this source for any purpose, but
reproduction or distribution in any format is prohibited without written
permission. See provided URL for inquiries about permission.
Persistent DSpace Link