  4. Experimental Demonstration of the Lower Leg Trajectory Error Framework Using Physiological Data as Inputs

Experimental Demonstration of the Lower Leg Trajectory Error Framework Using Physiological Data as Inputs

Thumbnail Image
Download
Name

bio_143_03_031003.pdf

Description
Published version
Size

1.87 MB

Format

Adobe PDF

Checksum (MD5)

b06ec99d4c06d2cc6c3166be9553de36

sword-2022-01-26T19:20:39.original.xml (130 B)
Original SWORD entry document
Author(s)
Olesnavage, Kathryn M
•
Prost, Victor
•
Johnson, William Brett
•
Major, Matthew J
•
Winter, Amos G
Date Issued
2021
Journal
Journal of Biomechanical Engineering
Publisher
ASME International
Citation
Olesnavage, Kathryn M, Prost, Victor, Johnson, William Brett, Major, Matthew J and Winter, Amos G. 2021. "Experimental Demonstration of the Lower Leg Trajectory Error Framework Using Physiological Data as Inputs." Journal of Biomechanical Engineering, 143 (3).
Version
Final published version
Abstract
© 2021 by ASME. While many studies have attempted to characterize the mechanical behavior of passive prosthetic feet to understand their influence on amputee gait, the relationship between mechanical design and biomechanical performance has not yet been fully articulated from a fundamental physics perspective. A novel framework, called lower leg trajectory error (LLTE) framework, presents a means of quantitatively optimizing the constitutive model of prosthetic feet to match a reference kinematic and kinetic dataset. This framework can be used to predict the required stiffness and geometry of a prosthesis to yield a desired biomechanical response. A passive prototype foot with adjustable ankle stiffness was tested by a unilateral transtibial amputee to evaluate this framework. The foot condition with LLTE-optimal ankle stiffness enabled the user to replicate the physiological target dataset within 16% root-mean-square (RMS) error. Specifically, the measured kinematic variables matched the target kinematics within 4% RMS error. Testing a range of ankle stiffness conditions from 1.5 to 24.4 N·m/deg with the same user indicated that conditions with lower LLTE values deviated the least from the target kinematic data. Across all conditions, the framework predicted the horizontal/vertical position, and angular orientation of the lower leg during midstance within 1.0 cm, 0.3 cm, and 1.5 deg, respectively. This initial testing suggests that prosthetic feet designed with low LLTE values could offer benefits to users. The LLTE framework is agnostic to specific foot designs and kinematic/kinetic user targets, and could be used to design and customize prosthetic feet.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
https://hdl.handle.net/1721.1/139754
DOI of Published Version
https://doi.org/10.1115/1.4048643
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.