Repository logo
Log in(current)
Repository logoMIT Open ScholarshipDSpace@MIT
  1. Home
  2. MIT Open Access Articles
  3. MIT Open Access Articles
  4. Comparison of ultrafast wave-controlled aliasing in parallel imaging (CAIPI) magnetization-prepared rapid acquisition gradient echo (MP-RAGE) and standard MP-RAGE in non-sedated children: initial clinical experience

Comparison of ultrafast wave-controlled aliasing in parallel imaging (CAIPI) magnetization-prepared rapid acquisition gradient echo (MP-RAGE) and standard MP-RAGE in non-sedated children: initial clinical experience

Thumbnail Image
Download
Name

247_2021_5117_ReferencePDF.pdf

Size

2.02 MB

Format

Adobe PDF

Checksum (MD5)

160631b5b614e8a613b601d0ec43a8d7

Download all files submitted through automated deposit
art_7438723463266524107.zip (1.93 MB)
Author(s)
Tabari, Azadeh
•
Conklin, John
•
Figueiro Longo, Maria G.
•
Jaimes, Camilo
•
Setsompop, Kawin
•
Cauley, Stephen F.
•
Kirsch, John E.
•
Huang, Susie Y.
•
Rapalino, Otto
•
Gee, Michael S.
more
Date Issued
July 15, 2021
Publisher
Springer Berlin Heidelberg
Citation
Tabari, Azadeh, Conklin, John, Figueiro Longo, Maria G., Jaimes, Camilo, Setsompop, Kawin et al. 2021. "Comparison of ultrafast wave-controlled aliasing in parallel imaging (CAIPI) magnetization-prepared rapid acquisition gradient echo (MP-RAGE) and standard MP-RAGE in non-sedated children: initial clinical experience."
Version
Author's final manuscript
Abstract
Abstract Background Fast magnetic resonance imaging (MRI) sequences are advantageous in pediatric imaging as they can lessen child discomfort, decrease motion artifact and improve scanner availability. Objective To evaluate the feasibility of an ultrafast wave-CAIPI (controlled aliasing in parallel imaging) MP-RAGE (magnetization-prepared rapid gradient echo) sequence for brain imaging of awake pediatric patients. Materials and methods Each MRI included a standard MP-RAGE sequence and an ultrafast wave-MP-RAGE sequence. Two neuroradiologists evaluated both sequences in terms of artifacts, noise, anatomical contrast and pathological contrast. A predefined 5-point scale was used by two independent pediatric neuroradiologists. A Wilcoxon signed-rank test was used to evaluate the difference between sequences for each variable. Results Twenty-four patients (14 males; mean age: 11.5±4.5 years, range: 1 month to 17.8 years) were included. Wave-CAIPI MP-RAGE provided a 77% reduction in scan time using a 32-channel coil and a 70% reduction using a 20-channel coil. Visualization of the pathology, artifacts and pathological enhancement (including parenchymal, leptomeningeal and dural enhancement) was not significantly different between standard MP-RAGE and wave-CAIPI MP-RAGE (all P>0.05). For central (P<0.001) and peripheral (P<0.001) noise, and the evaluation of the anatomical structures (P<0.001), the observers favored standard MP-RAGE over wave-CAIPI MP-RAGE. Conclusion Ultrafast brain imaging with wave-CAIPI MP-RAGE is feasible in awake pediatric patients, providing a substantial reduction in scan time at a cost of subjectively increased image noise.
MIT Department
Harvard University--MIT Division of Health Sciences and Technology
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
https://hdl.handle.net/1721.1/136742
DOI of Published Version
https://doi.org/10.1007/s00247-021-05117-5
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.