A miniature, implantable wireless neural stimulation system
Name
144553703-MIT.pdf
Description
Full printable version
Size
7.88 MB
Format
Adobe PDF
Checksum (MD5)
349a7e5b34c0a7ed8a8836fbfaba4f64
Author(s)
Arfin, Scott K. (Scott Kenneth)
Advisor(s)
Rahul Sarpeshkar.
Date Issued
2006
Publisher
Massachusetts Institute of Technology
Abstract
In this thesis, I present the design of a wireless neural stimulation system. The system consists of an external transmitter, controllable through a computer interface, and a miniature, implantable wireless receiver and stimulator. The implant is tailored for use in zebra finches - small birds weighing just 12-15g - as part of ongoing research into the neural mechanisms of sequence generation and learning. The implant, assembled on a miniature printed circuit board, contains a receiver coil, battery, electrodes, and a custom integrated circuit for data demodulation and neural stimulation. The chip, fabricated in a standard 0.5[mu]m CMOS process, is capable of delivering biphasic current pulses to 4 addressable electrode sites at 16 selectable current levels ranging from 100[mu]A to mA. Additionally, the biphasic pulses may be inverted. The entire implant weighs less than 1.5g and occupies a footprint smaller than 1.5cm2. A miniaturized neural stimulator such as this one also has applications in neural prostheses for blindness, Parkinson's disease, and paralysis.
Description
Thesis (S.M.)--Massachusetts Institute of Technology, Dept. of Electrical Engineering and Computer Science, 2006.
Includes bibliographical references (p. 103-105).
Subjects
Electrical Engineering and Computer Science.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Terms of Use
M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.