May the force be with you : a medical ultrasound system with integrated force measurement
Name
1021887214-MIT.pdf
Description
Full printable version
Size
16.2 MB
Format
Adobe PDF
Checksum (MD5)
ebdce72270626a0f85edb40ac45e0a61
Author(s)
Huang, Athena Yeh
Advisor(s)
Brian W. Anthony.
Alternative Title
Medical ultrasound system with integrated force measurement
Date Issued
2017
Publisher
Massachusetts Institute of Technology
Abstract
Ultrasound is a commonly used medical imaging modality for non-invasive examination of soft tissue. Clinical ultrasound scanning requires significant contact of the transducer face with the patient, and the contact force exerted by the sonographer can vary widely throughout a scan or from scan to scan. This thesis explores the design and development of an electromechanical system to measure the contact force during ultrasound scanning. The device is a handheld ultrasound probe with force sensors integrated into its housing such that the force distribution across the ultrasound transducer face can be measured. The device was used to perform shear wave elastography on tissue-mimicking phantoms and ex vivo tissue at varying force distributions. A gradient in applied pressure introduced a gradient in elasticity across the image for ex vivo tissue but not for phantoms. To consider how the device integrates into the overall system, a human factors study was done to compare feedback modalities provided to the human operator during ultrasound scanning. Visual feedback was more effective than haptic feedback for force tracking, but at expense of ultrasound path tracking. Lastly, two methods of data synchronization of the acquired force data and ultrasound images are considered. A three tap software synchronization method is a feasible alternative when hardware synchronization is unavailable. As a whole, this system will improve the repeatability and capabilities of ultrasound imaging.
Description
Thesis: S.M., Massachusetts Institute of Technology, Department of Mechanical Engineering, 2017.
Cataloged from PDF version of thesis.
Includes bibliographical references (pages 107-110).
Subjects
Mechanical Engineering.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
MIT theses are protected by copyright. They may be viewed, downloaded, or printed from this source but further reproduction or distribution in any format is prohibited without written permission.