Quantitative cerebral blood flow with optical coherence tomography
Name
Fujimoto-Quantitative Cerebral.pdf
Size
4.86 MB
Format
Adobe PDF
Checksum (MD5)
b933f9aa93a3f36fc1b1b4d42e0685e2
Author(s) • • • • • •
Srinivasan, Vivek J.
Sakadžic, Sava
Gorczynska, Iwona
Ruvinskaya, Svetlana
Wu, Weicheng
Fujimoto, James G.
Boas, David A.
Date Issued
January 2010
Journal
Optics Express
Publisher
Optical Society of America
Citation
Srinivasan, Vivek J. et al. “Quantitative Cerebral Blood Flow with Optical Coherence Tomography.” Optics Express 18.3 (2010): 2477.
Version
Author's final manuscript
Abstract
Absolute measurements of cerebral blood flow (CBF) are an important endpoint in studies of cerebral pathophysiology. Currently no accepted method exists for in vivo longitudinal monitoring of CBF with high resolution in rats and mice. Using three-dimensional Doppler Optical Coherence Tomography and cranial window preparations, we present methods and algorithms for regional CBF measurements in the rat cortex. Towards this end, we develop and validate a quantitative statistical model to describe the effect of static tissue on velocity sensitivity. This model is used to design scanning protocols and algorithms for sensitive 3D flow measurements and angiography of the cortex. We also introduce a method of absolute flow calculation that does not require explicit knowledge of vessel angles. We show that OCT estimates of absolute CBF values in rats agree with prior measures by autoradiography, suggesting that Doppler OCT can perform absolute flow measurements in animal models.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Massachusetts Institute of Technology. Research Laboratory of Electronics
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike 3.0
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1364/OE.18.002477