Sensitivity of MEG and EEG to Source Orientation
Name
Ahlfors_Sensitivity of.pdf
Size
484.4 KB
Format
Adobe PDF
Checksum (MD5)
ec676df72752187e98d0f86aa111bc1c
Author(s) • • •
Ahlfors, Seppo Pentti
Han, Jooman
Belliveau, John W.
Hamalainen, Matti S.
Date Issued
July 2010
Journal
Brain Topography
Publisher
Springer
Citation
Ahlfors, Seppo P. et al. “Sensitivity of MEG and EEG to Source Orientation.” Brain Topography 23.3 (2010) : 227-232. Copyright © 2010, The Author(s)
Version
Final published version
Abstract
An important difference between magnetoencephalography
(MEG) and electroencephalography (EEG)
is that MEG is insensitive to radially oriented sources. We
quantified computationally the dependency of MEG and
EEG on the source orientation using a forward model with
realistic tissue boundaries. Similar to the simpler case of a
spherical head model, in which MEG cannot see radial
sources at all, for most cortical locations there was a source
orientation to which MEG was insensitive. The median
value for the ratio of the signal magnitude for the source
orientation of the lowest and the highest sensitivity was
0.06 for MEG and 0.63 for EEG. The difference in the
sensitivity to the source orientation is expected to contribute
to systematic differences in the signal-to-noise ratio
between MEG and EEG.
MIT Department
Harvard University--MIT Division of Health Sciences and Technology
Terms of Use
Creative Commons Attribution Noncommercial License
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1007/s10548-010-0154-x