The CUORE Cryostat: A 1-Ton Scale Setup for Bolometric Detectors
Name
10909_2015_Article_1389.pdf
Size
751.2 KB
Format
Adobe PDF
Checksum (MD5)
fd275fc33c5180623a301f45c090d5af
Author(s) • • • • • • • • •
Ligi, C.
Alduino, C.
Alessandria, F.
Biassoni, M.
Bucci, C.
Caminata, A.
Canonica, L.
Cappelli, L.
Chott, N. I
Copello, S.
Date Issued
December 2015
Journal
Journal of Low Temperature Physics
Publisher
Springer US
Citation
Ligi, C. et al. “The CUORE Cryostat: A 1-Ton Scale Setup for Bolometric Detectors.” Journal of Low Temperature Physics 184.3–4 (2016): 590–596.
Version
Author's final manuscript
Abstract
The cryogenic underground observatory for rare events (CUORE) is a 1-ton scale bolometric experiment whose detector consists of an array of 988 TeO[subscript 2] crystals arranged in a cylindrical compact structure of 19 towers. This will be the largest bolometric mass ever operated. The experiment will work at a temperature around or below 10 mK. CUORE cryostat consists of a cryogen-free system based on pulse tubes and a custom high power dilution refrigerator, designed to match these specifications. The cryostat has been commissioned in 2014 at the Gran Sasso National Laboratories and reached a record temperature of 6 mK on a cubic meter scale. In this paper, we present results of CUORE commissioning runs. Details on the thermal characteristics and cryogenic performances of the system will be also given.
MIT Department
Massachusetts Institute of Technology. Department of Physics
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.