  4. Enumeration of CD4+ T-Cells Using a Portable Microchip Count Platform in Tanzanian HIV-Infected Patients

Enumeration of CD4+ T-Cells Using a Portable Microchip Count Platform in Tanzanian HIV-Infected Patients

Thumbnail Image
Download
Name

Moon-2011-Enumeration of CD4(+.pdf

Size

1.11 MB

Format

Adobe PDF

Checksum (MD5)

efeab572a4546251c18d2944904e4eee

Author(s)
Moon, SangJun
•
Gurkan, Umut Atakan
•
Blander, Jeffrey
•
Fawzi, Wafaie W.
•
Aboud, Said
•
Mugusi, Ferdinand
•
Kuritzkes, Daniel R.
•
Demirci, Utkan
Date Issued
April 2011
Journal
PLoS ONE
Publisher
Public Library of Science
Citation
Moon, SangJun, et al. (2011) "Enumeration of CD4+ T-Cells Using a Portable Microchip Count Platform in Tanzanian HIV-Infected Patients." PLoS ONE 6(7): e21409.
Version
Final published version
Abstract
Background CD4+ T-lymphocyte count (CD4 count) is a standard method used to monitor HIV-infected patients during anti-retroviral therapy (ART). The World Health Organization (WHO) has pointed out or recommended that a handheld, point-of-care, reliable, and affordable CD4 count platform is urgently needed in resource-scarce settings. Methods HIV-infected patient blood samples were tested at the point-of-care using a portable and label-free microchip CD4 count platform that we have developed. A total of 130 HIV-infected patient samples were collected that included 16 de-identified left over blood samples from Brigham and Women's Hospital (BWH), and 114 left over samples from Muhimbili University of Health and Allied Sciences (MUHAS) enrolled in the HIV and AIDS care and treatment centers in the City of Dar es Salaam, Tanzania. The two data groups from BWH and MUHAS were analyzed and compared to the commonly accepted CD4 count reference method (FACSCalibur system). Results The portable, battery operated and microscope-free microchip platform developed in our laboratory (BWH) showed significant correlation in CD4 counts compared with FACSCalibur system both at BWH (r = 0.94, p<0.01) and MUHAS (r = 0.49, p<0.01), which was supported by the Bland-Altman methods comparison analysis. The device rapidly produced CD4 count within 10 minutes using an in-house developed automated cell counting program. Conclusions We obtained CD4 counts of HIV-infected patients using a portable platform which is an inexpensive (<$1 material cost) and disposable microchip that uses whole blood sample (<10 µl) without any pre-processing. The system operates without the need for antibody-based fluorescent labeling and expensive fluorescent illumination and microscope setup. This portable CD4 count platform displays agreement with the FACSCalibur results and has the potential to expand access to HIV and AIDS monitoring using fingerprick volume of whole blood and helping people who suffer from HIV and AIDS in resource-limited settings.
MIT Department
Harvard University--MIT Division of Health Sciences and Technology
Terms of Use
Creative Commons Attribution
http://creativecommons.org/licenses/by/2.5/
Persistent DSpace Link
http://hdl.handle.net/1721.1/66273
DOI of Published Version
https://doi.org/10.1371/journal.pone.0021409
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.