Cardiac Ablation Catheter Guidance by Means of a Single Equivalent Moving Dipole Inverse Algorithm
Name
Cohen_Cardiac ablation.pdf
Size
1.71 MB
Format
Adobe PDF
Checksum (MD5)
182f6871497a827925a0d93fa2d55172
Author(s) • • • • • • • • •
Lee, Kichang
Lv, Wener
Ter-Ovanesyan, Evgeny
Barley, Maya E.
Voysey, Graham E.
Galea, Anna M.
Hirschman, Gordon B.
Leroy, Kristen
Marini, Robert P.
Barrett, Conor
Date Issued
February 2013
Journal
Pacing and Clinical Electrophysiology
Publisher
Wiley Blackwell
Citation
Lee, Kichang, Wener Lv, Evgeny Ter-Ovanesyan, Maya E. Barley, Graham E. Voysey, Anna M. Galea, Gordon B. Hirschman, et al. “Cardiac Ablation Catheter Guidance by Means of a Single Equivalent Moving Dipole Inverse Algorithm.” Pacing and Clinical Electrophysiology 36, no. 7 (July 2013): 811–22.
Version
Author's final manuscript
Abstract
Background
We developed and evaluated a novel system for guiding radiofrequency catheter ablation therapy of ventricular tachycardia. This guidance system employs an inverse solution guidance algorithm (ISGA) using a single equivalent moving dipole (SEMD) localization method. The method and system were evaluated in both a saline tank phantom model and in vivo animal (swine) experiments.
Methods
A catheter with two platinum electrodes spaced 3 mm apart was used as the dipole source in the phantom study. A 40-Hz sinusoidal signal was applied to the electrode pair. In the animal study, four to eight electrodes were sutured onto the right ventricle. These electrodes were connected to a stimulus generator delivering 1-ms duration pacing pulses. Signals were recorded from 64 electrodes, located either on the inner surface of the saline tank or on the body surface of the pig, and then processed by the ISGA to localize the physical or bioelectrical SEMD.
Results
In the phantom studies, the guidance algorithm was used to advance a catheter tip to the location of the source dipole. The distance from the final position of the catheter tip to the position of the target dipole was 2.22 ± 0.78 mm in real space and 1.38 ± 0.78 mm in image space (computational space). The ISGA successfully tracked the locations of electrodes sutured on the ventricular myocardium and the movement of an endocardial catheter placed in the animal's right ventricle.
Conclusion
In conclusion, we successfully demonstrated the feasibility of using an SEMD inverse algorithm to guide a cardiac ablation catheter.
MIT Department
Institute for Medical Engineering and Science
Harvard University--MIT Division of Health Sciences and Technology
Massachusetts Institute of Technology. Department of Mechanical Engineering
Massachusetts Institute of Technology. Division of Comparative Medicine
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1111/pace.12114