Titanium Coating of the Boston Keratoprosthesis
Name
Salvador-Culla-2016-Titanium Coating of.pdf
Size
1.14 MB
Format
Adobe PDF
Checksum (MD5)
7bb1c332c6b0b6d769c3b967211b4a70
Author(s) • • • • • •
Jeong, Kyung Jae
Kolovou, Paraskevi Evi
Chodosh, James
Dohlman, Claes H.
Kohane, Daniel S.
Chiang, Homer H.
Salvador Culla, Borja
Date Issued
April 2016
Journal
Translational Vision Science & Technology
Publisher
Association for Research in Vision and Ophthalmology
Citation
Salvador-Culla, Borja et al. “Titanium Coating of the Boston Keratoprosthesis.” Translational Vision Science & Technology 5.2 (2016): 17. © 2015 Association for Research in Vision and Ophthalmology
Version
Final published version
Abstract
Purpose: We tested the feasibility of using titanium to enhance adhesion of the Boston Keratoprosthesis (B-KPro), ultimately to decrease the risk of implant-associated complications. Methods: Cylindrical rods were made of poly(methyl methacrylate) (PMMA), PMMA coated with titanium dioxide (TiO[subscript 2]) over a layer of polydopamine (PMMA[subscript TiO2]), smooth (Ti) and sandblasted (Ti[subscript SB]) titanium, and titanium treated with oxygen plasma (Ti[subscript ox] and Ti[subscript SBox]). Topography and surface chemistry were analyzed by scanning electron microscopy (SEM), atomic force microscopy (AFM), and X-ray photoelectron spectroscopy (XPS). Adhesion force between rods and porcine corneas was measured ex vivo. Titanium sleeves, smooth and sandblasted, were inserted around the stem of the B-KPro and implanted in rabbits. Tissue adhesion to the stem was assessed and compared to an unmodified B-Kpro after 1 month. Results: X-ray photoelectron spectroscopy demonstrated successful deposition of TiO[subscript 2] on polydopamine-coated PMMA. Oxygen plasma treatment did not change the
XPS spectra of titanium rods (Ti and Ti[subscript SB]), although it increased their hydrophilicity. The materials did not show cell toxicity. After 14 days of incubation, PMMA[subscript TiO2], smooth titanium treated with oxygen plasma (Ti[subscript ox]), and sandblasted titanium rods (Ti[subscript SB], Ti[subscript SBox]) showed significantly higher adhesion forces than PMMA ex vivo. In vivo, the use of a Ti[subscript SB] sleeve around the stem of the B-KPro induced a significant increase in tissue adhesion compared to a Ti sleeve or bare PMMA. Conclusions: Sandblasted titanium sleeves greatly enhanced adherence of the B-KPro to the rabbit cornea. This approach may improve adhesion with the donor cornea in humans as well. Translational Relevance: This approach may improve adhesion with donor corneas in humans.
MIT Department
Koch Institute for Integrative Cancer Research at MIT
Terms of Use
Creative Commons Attribution-NonCommercial-NoDerivs License
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1167/tvst.5.2.17