Repository logo
Log in(current)
Repository logoMIT Open ScholarshipDSpace@MIT
  1. Home
  2. MIT Open Access Articles
  3. MIT Open Access Articles
  4. Simplified Negative Pressure Wound Therapy Device for Application in Low-Resource Settings

Simplified Negative Pressure Wound Therapy Device for Application in Low-Resource Settings

Thumbnail Image
Download
Name

nihms710509.pdf

Size

287.12 KB

Format

Adobe PDF

Checksum (MD5)

cafce3b8858c9d5873738ef0b145e2ba

Download all files submitted through automated deposit
sword-2019-01-02.original.zip (234.92 KB)
Author(s)
Zurovcik, Danielle R.
•
Mody, Gita N.
•
Riviello, Robert
•
Slocum, Alexander H
Date Issued
October 2015
Journal
Journal of Orthopaedic Trauma
Publisher
Ovid Technologies (Wolters Kluwer Health)
Citation
Zurovcik, Danielle R., Gita N. Mody, Robert Riviello, and Alex Slocum. “Simplified Negative Pressure Wound Therapy Device for Application in Low-Resource Settings.” Journal of Orthopaedic Trauma 29 (October 2015): S33–S36.
Version
Author's final manuscript
Abstract
Negative pressure wound therapy (NPWT) provides proven wound healing benefits and is often a desirable wound treatment methodology. Unfortunately, NPWT devices are not widely available in low-resource settings. To overcome the identified NPWT barriers, a simplified NPWT (sNPWT) system was designed and iteratively improved during field-based testing. The sNPWT technology, our device design iterations, and the design-based results of our field tests are described in this article. The sNPWT system includes a bellows hand pump, an occlusive drape, and a tube with tube connectors, connecting the drape to the pump. The most critical property of an sNPWT system is that it must be airtight. The details of the design iterations, which are needed to achieve an occlusive system, are explained. During the design process, the sNPWT system was tested during the earthquake relief in Haiti. This testing found that a liquid sealant was necessary to seal the drape to the periwound skin. A study conducted in Rwanda verified that a liquid latex sealant was safe to use, and that the tube connector must be connected to the drape with an airtight method during the manufacturing process. This work has shown that sNPWT is feasible in low-resource settings. Since the completion of the clinical testing, the design has been further evolved, and the developers are working with contract manufacturers to produce the final design and preparing for regulatory approval applications.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
http://creativecommons.org/licenses/by-nc-sa/4.0/
Persistent DSpace Link
http://hdl.handle.net/1721.1/120130
DOI of Published Version
https://doi.org/10.1097/BOT.0000000000000410
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.