  4. Near-common-path interferometer for imaging Fourier-transform spectroscopy in wide-field microscopy

Near-common-path interferometer for imaging Fourier-transform spectroscopy in wide-field microscopy

Thumbnail Image
Download
Name

nihms925583.pdf

Size

1.43 MB

Format

Adobe PDF

Checksum (MD5)

3f90da27a56fb20b241cce0e291ddbbf

Download all files submitted through automated deposit
sword-2019-01-04.original.zip (1.17 MB)
Author(s)
Heemskerk, Hans
•
Wadduwage, Dushan Nawoda
•
Singh, Vijay Raj
•
Choi, Heejin
•
Yaqoob, Zahid
•
So, Peter T. C.
•
Matsudaira, Paul T.
Date Issued
May 2017
Journal
Optica
Publisher
The Optical Society
Citation
Wadduwage, Dushan N., Vijay Raj Singh, Heejin Choi, Zahid Yaqoob, Hans Heemskerk, Paul Matsudaira, and Peter T. C. So. “Near-Common-Path Interferometer for Imaging Fourier-Transform Spectroscopy in Wide-Field Microscopy.” Optica 4, no. 5 (May 16, 2017): 546.
Version
Author's final manuscript
Abstract
Imaging Fourier-transform spectroscopy (IFTS) is a powerful method for biological hyperspectral analysis based on various imaging modalities, such as fluorescence or Raman. Since the measurements are taken in the Fourier space of the spectrum, it can also take advantage of compressed sensing strategies. IFTS has been readily implemented in high-throughput, high-content microscope systems based on wide-field imaging modalities. However, there are limitations in existing wide-field IFTS designs. Non-common-path approaches are less phase-stable. Alternatively, designs based on the common-path Sagnac interferometer are stable, but incompatible with high-throughput imaging. They require exhaustive sequential scanning over large interferometric path delays, making compressive strategic data acquisition impossible. In this paper, we present a novel phase-stable, near-common-path interferometer enabling high-throughput hyperspectral imaging based on strategic data acquisition. Our results suggest that this approach can improve throughput over those of many other wide-field spectral techniques by more than an order of magnitude without compromising phase stability.
MIT Department
Massachusetts Institute of Technology. Institute for Medical Engineering & Science
Massachusetts Institute of Technology. Department of Biological Engineering
Massachusetts Institute of Technology. Department of Biology
Massachusetts Institute of Technology. Department of Chemistry
Massachusetts Institute of Technology. Department of Mechanical Engineering
Massachusetts Institute of Technology. Spectroscopy Laboratory
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
http://creativecommons.org/licenses/by-nc-sa/4.0/
Persistent DSpace Link
http://hdl.handle.net/1721.1/120327
DOI of Published Version
https://doi.org/10.1364/OPTICA.4.000546