Leveraging device-arterial coupling to determine cardiac and vascular state
Name
nihms-1024477.pdf
Description
Accepted version
Size
435.53 KB
Format
Adobe PDF
Checksum (MD5)
068e19c6dce473c0442c0d17ca4abb19
Author(s) • •
Chang, Brian Yale
Keller, Steven
Edelman, Elazer R
Date Issued
October 2019
Journal
IEEE Transactions on Biomedical Engineering
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Chang, Brian Y., Steven P. Keller and Elazer R. Edelman. “Leveraging device-arterial coupling to determine cardiac and vascular state.” IEEE Transactions on Biomedical Engineering, 66, 10 (October 2019): 2800-2808 © 2019 The Author(s)
Version
Author's final manuscript
Abstract
Limitations in available diagnostic metrics restrict the efficacy of managing therapies for cardiogenic shock. In current clinical practice, cardiovascular state is inferred through measurement of pulmonary capillary wedge pressure and reliance on linear approximations between pressure and flow to estimate peripheral vascular resistance. Mechanical circulatory support devices residing within the left ventricle and aorta provide an opportunity for both determining cardiac and vascular state and offering therapeutic benefit. We leverage the controllable mode of operation and transvalvular position of an indwelling percutaneous ventricular assist device to assess vascular and, in turn, cardiac state through the effects of device-arterial coupling across different levels of device support. Methods: Vascular state is determined by measuring changes in the pressure waveforms induced through intentional variation in the device generated blood flow. We evaluate this impact by applying a lumped parameter model to quantify state-specific vascular resistance and compliance and calculate beat-to-beat stroke volume and cardiac output in both animal models and retrospective patient data without external calibration. Results: Vascular state was accurately predicted in patients and animals in both baseline and experimental conditions. In the animal, stroke volume was predicted within a total root mean square error of 3.71 mL (n = 482). Conclusion: We demonstrate that device-arterial coupling is a powerful tool for evaluating patient and state specific parameters of cardiovascular function. Significance: These insights may yield improved clinical care and support the development of next generation mechanical circulatory support devices that determine and operate in tandem with the supported organ.
MIT Department
Massachusetts Institute of Technology. Institute for Medical Engineering & Science
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1109/TBME.2019.2895752