Abstract
The development of scalable, high-precision measurement systems is a critical enabling step for both preclinical drug testing using engineered muscle tissue and the emerging field of biohybrid actuation. This thesis investigates the mechanical, sensing, and manufacturing principles required to accurately quantify contractile behavior, with particular focus on variable stiffness, controllable tension, and integrated strain sensing within a compact, biocompatible device.
A flexure-based architecture was developed to provide (i) discrete and continuous modulation of environmental stiffness over the range required for muscle growth and contraction, (ii) automated adjustment of muscle preload, and (iii) direct strain measurement using custom gauges electroplated onto a low-density polyethylene (LDPE) substrate. The system was designed to operate within a standard 24-well plate, enabling scalable testing of hiPSC-derived skeletal muscle tissues.
Experimental results showed that variable boundary conditions could shift the effective stiffness experienced by a single muscle sample from 0.38 \pm\ 0.04 N/m to 3.85 \pm 0.11 N/m without repositioning the muscle, while motorized tensioning achieved controlled pre-strain from 0-20%. Integrated strain gauges resolved micron-scale displacements and dynamic contractions, providing an analog alternative to image-based methods.
Fabrication studies revealed relationships between polymer compliance, thermal expansion, and metallization integrity. Temperature gradients, part size, metal thickness, and process sequence were found to govern residual stress and cracking, establishing practical design rules for plating on soft substrates. Stress-relief conditioning and reduced characteristic length significantly improved plating quality, and preliminary results suggest that lower-temperature, more ductile copper plating may further reduce failure risk relative to nickel.
Together, these results define key principles for designing scalable, automated, and mechanically tunable platforms for high-precision measurement of engineered muscle and future biohybrid actuators.