A Novel Lead Garment Structural System to Alleviate Orthopedic Stress for Surgeons
Name
v001t09a002-dmd2018-6920.pdf
Size
1.54 MB
Format
Adobe PDF
Checksum (MD5)
04d8477c5c1261516efd0f4861cb7b13
Author(s) • • • • • • •
Johnson, Hilary Anna
Miller, Sally A.
Tawde, Prianca
LaPenta, Bethany Anne
Teo, Daniel
Cervantes, Thomas Michael
Mehta, Nishaki
Slocum, Alexander H
Date Issued
April 2018
Journal
2018 Design of Medical Devices Conference
Publisher
American Society of Mechanical Engineers (ASME)
Citation
Johnson, Hilary et al. “A Novel Lead Garment Structural System to Alleviate Orthopedic Stress for Surgeons.” 2018 Design of Medical Devices Conference, April 2018, Minneapolis, Minnesota, USA, American Society of Mechanical Engineers, April 2018 © American Society of Mechanical Engineers
Version
Final published version
Abstract
Cardiovascular, orthopedic, and interventional radiology procedures using fluoroscopy require healthcare professionals to wear heavy lead garments for radiation protection, sometimes for up to 12 hours per day. Wearing lead garments for prolonged periods of time can lead to musculoskeletal injuries, discomfort, and fatigue. MobiLead is a mobile lead garment frame that was developed to reduce the weight supported by the user in an effort to mitigate these problems. The MobiLead system moves the lower garment load off the user’s body to a structural ground-supported frame and redistributes the upper load from the shoulders to the hips through a torso frame. The system is compact and maximizes the limited space available in operating rooms, while still giving the surgeon adequate mobility for various emergency procedures. Preliminary analysis of device effectiveness was conducted using electromyography and qualitative surgeon user feedback surveys. This paper will discuss the design, fabrication, and testing procedures for this mobile radiation protection system optimizing both support and mobility.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Terms of Use
Article is made available in accordance with the publisher's policy and may be subject to US copyright law. Please refer to the publisher's site for terms of use.
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1115/DMD2018-6920