Repository logo
Log in(current)
Repository logoMIT Open ScholarshipDSpace@MIT
  1. Home
  2. MIT Open Access Articles
  3. MIT Open Access Articles
  4. A comparison of point-tracking algorithms in ultrasound videos from the upper limb

A comparison of point-tracking algorithms in ultrasound videos from the upper limb

Thumbnail Image
Download
Name

12938_2023_Article_1105.pdf

Size

3.06 MB

Format

Adobe PDF

Checksum (MD5)

f2959e813e471766eb9dfa9820d4b0cd

Download all files submitted through automated deposit
art_4870505772328097782.zip (2.71 MB)
Author(s)
Magana-Salgado, Uriel
•
Namburi, Praneeth
•
Feigin-Almon, Micha
•
Pallares-Lopez, Roger
•
Anthony, Brian
Date Issued
May 24, 2023
Publisher
BioMed Central
Citation
BioMedical Engineering OnLine. 2023 May 24;22(1):52
Version
Final published version
Abstract
Abstract Tracking points in ultrasound (US) videos can be especially useful to characterize tissues in motion. Tracking algorithms that analyze successive video frames, such as variations of Optical Flow and Lucas–Kanade (LK), exploit frame-to-frame temporal information to track regions of interest. In contrast, convolutional neural-network (CNN) models process each video frame independently of neighboring frames. In this paper, we show that frame-to-frame trackers accumulate error over time. We propose three interpolation-like methods to combat error accumulation and show that all three methods reduce tracking errors in frame-to-frame trackers. On the neural-network end, we show that a CNN-based tracker, DeepLabCut (DLC), outperforms all four frame-to-frame trackers when tracking tissues in motion. DLC is more accurate than the frame-to-frame trackers and less sensitive to variations in types of tissue movement. The only caveat found with DLC comes from its non-temporal tracking strategy, leading to jitter between consecutive frames. Overall, when tracking points in videos of moving tissue, we recommend using DLC when prioritizing accuracy and robustness across movements in videos, and using LK with the proposed error-correction methods for small movements when tracking jitter is unacceptable.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Massachusetts Institute of Technology. Institute for Medical Engineering & Science
Terms of Use
Creative Commons Attribution
http://creativecommons.org/licenses/by/4.0/
Persistent DSpace Link
https://hdl.handle.net/1721.1/150827
DOI of Published Version
https://doi.org/10.1186/s12938-023-01105-y