A robot for gait rehabilitation
Name
71150008-MIT.pdf
Description
Full printable version
Size
17.46 MB
Format
Adobe PDF
Checksum (MD5)
c3bf35a623fb5ea0e727325a6bc14d38
Author(s)
Roberts, Michael (Michael Henry), 1980-
Advisor(s)
Herman Igo Krebs.
Date Issued
2004
Publisher
Massachusetts Institute of Technology
Abstract
After a stroke, persons suffer from neurological impairments that affect gait, and so require rehabilitation to regain ambulatory function. While 82% of patients recover the ability to walk, current methods including physiotherapy and partial body-weight supported treadmill training (PBWSTT) are monotonous and require intense therapist effort. The Mechanized Gait Trainer and the LOKOMAT are two robotic devices that have been developed to improve gait rehabilitation, but neither provides the facilitation of pelvis movements afforded by traditional methods. In addition, neither device is truly backdrivable. As shown by Hogan and Krebs, backdrivable, impedance-controlled robots are ideal for rehabilitation because of their stable interaction properties. Robots for the arm/shoulder, wrist, ankle, and hand have already been developed. This thesis describes the design of a robot for gait rehabilitation through the facilitation of pelvis movements. Four degrees of freedom (DOF) are actuated: vertical, lateral, and frontal translations as well as the rotation about the vertical axis. Vertical forces support part of the patient's weight. Lateral forces assist the weight shift from stance leg to swing leg and are a part of physiotherapy and treadmill training.
(cont.) Frontal forces help pace the subject as on a treadmill. Pelvic rotations can impart energy into the swing leg without direct actuation of the hip and knee muscles. A four DOF mechanism was designed to control these movements, consisting of a three-DOF planar linkage with a vertical prismatic translation. A mockup of the configuration was designed and tested to show that the non-actuated pelvis DOFs are not adversely affected by the device. Design calculations include finding the optimal linkage configuration, selecting ballspline shafts for the vertical DOF, selecting actuators, and designing the robot arm cross-sections and joints. A final design for the four-DOF module is presented.
Description
Thesis (S.M.)--Massachusetts Institute of Technology, Dept. of Mechanical Engineering, 2004.
Includes bibliographical references (p. 216-220).
Subjects
Mechanical Engineering.
MIT Department
Massachusetts Institute of Technology. Department of Mechanical Engineering
Terms of Use
M.I.T. theses are protected by copyright. They may be viewed from this source for any purpose, but reproduction or distribution in any format is prohibited without written permission. See provided URL for inquiries about permission.
Persistent DSpace Link