Local SAR, global SAR, transmitter power and excitation accuracy trade-offs in low flip-angle parallel transmit pulse design
Name
Local specific.pdf
Size
3.58 MB
Format
Adobe PDF
Checksum (MD5)
1af9716dbdfa93ad5c0e547be9592366
Author(s) • • • •
Guérin, Bastien
Gebhardt, Matthias
Cauley, Steven
Adalsteinsson, Elfar
Wald, Lawrence L.
Alternative Title
Local specific absorption rate (SAR), global SAR, transmitter power, and excitation accuracy trade-offs in low flip-angle parallel transmit pulse design
Date Issued
June 2013
Journal
Magnetic Resonance in Medicine
Publisher
Wiley Blackwell
Citation
Guérin, Bastien, Matthias Gebhardt, Steven Cauley, Elfar Adalsteinsson, and Lawrence L. Wald. “Local Specific Absorption Rate (SAR), Global SAR, Transmitter Power, and Excitation Accuracy Trade-Offs in Low Flip-Angle Parallel Transmit Pulse Design.” Magn. Reson. Med. 71, no. 4 (June 14, 2013): 1446–1457.
Version
Author's final manuscript
Abstract
Purpose
We propose a constrained optimization approach for designing parallel transmit (pTx) pulses satisfying all regulatory and hardware limits. We study the trade-offs between excitation accuracy, local and global specific absorption rate (SAR), and maximum and average power for small flip-angle pTx (eight channels) spokes pulses in the torso at 3 T and in the head at 7 T.
Methods
We compare the trade-offs between the above-mentioned quantities using the L-curve method. We use a primal-dual algorithm and a compressed set of local SAR matrices to design radio-frequency (RF) pulses satisfying all regulatory (including local SAR) and hardware constraints.
Results
Local SAR can be substantially reduced (factor of 2 or more) by explicitly constraining it in the pulse design process compared to constraining global SAR or pulse power alone. This often comes at the price of increased pulse power.
Conclusion
Simultaneous control of power and SAR is needed for the design of pTx pulses that are safe and can be played on the scanner. Constraining a single quantity can create large increase in the others, which can then rise above safety or hardware limits. Simultaneous constraint of local SAR and power is fast enough to be applicable in a clinical setting.