Robust control of burst suppression for medical coma
Name
Brown_Robust control.pdf
Size
2.3 MB
Format
Adobe PDF
Checksum (MD5)
98dfc1335b17fce59957e0f21217a3f6
Author(s) • • • •
Kim, Seong-Eun
Ching, ShiNung
Brown, Emery N.
Westover, M. Brandon
Purdon, Patrick L.
Date Issued
May 2015
Journal
Journal of Neural Engineering
Publisher
IOP Publishing
Citation
Westover, M Brandon, Seong-Eun Kim, ShiNung Ching, Patrick L Purdon, and Emery N Brown. “Robust Control of Burst Suppression for Medical Coma.” Journal of Neural Engineering 12, no. 4 (May 28, 2015): 046004.
Version
Author's final manuscript
Abstract
Objective. Medical coma is an anesthetic-induced state of brain inactivation, manifest in the electroencephalogram by burst suppression. Feedback control can be used to regulate burst suppression, however, previous designs have not been robust. Robust control design is critical under real-world operating conditions, subject to substantial pharmacokinetic and pharmacodynamic parameter uncertainty and unpredictable external disturbances. We sought to develop a robust closed-loop anesthesia delivery (CLAD) system to control medical coma. Approach. We developed a robust CLAD system to control the burst suppression probability (BSP). We developed a novel BSP tracking algorithm based on realistic models of propofol pharmacokinetics and pharmacodynamics. We also developed a practical method for estimating patient-specific pharmacodynamics parameters. Finally, we synthesized a robust proportional integral controller. Using a factorial design spanning patient age, mass, height, and gender, we tested whether the system performed within clinically acceptable limits. Throughout all experiments we subjected the system to disturbances, simulating treatment of refractory status epilepticus in a real-world intensive care unit environment. Main results. In 5400 simulations, CLAD behavior remained within specifications. Transient behavior after a step in target BSP from 0.2 to 0.8 exhibited a rise time (the median (min, max)) of 1.4 [1.1, 1.9] min; settling time, 7.8 [4.2, 9.0] min; and percent overshoot of 9.6 [2.3, 10.8]%. Under steady state conditions the CLAD system exhibited a median error of 0.1 [−0.5, 0.9]%; inaccuracy of 1.8 [0.9, 3.4]%; oscillation index of 1.8 [0.9, 3.4]%; and maximum instantaneous propofol dose of 4.3 [2.1, 10.5] mg kg[superscript −1]. The maximum hourly propofol dose was 4.3 [2.1, 10.3] mg kg[superscript −1] h[superscript −1]. Performance fell within clinically acceptable limits for all measures. Significance. A CLAD system designed using robust control theory achieves clinically acceptable performance in the presence of realistic unmodeled disturbances and in spite of realistic model uncertainty, while maintaining infusion rates within acceptable safety limits.
MIT Department
Massachusetts Institute of Technology. Department of Brain and Cognitive Sciences
Picower Institute for Learning and Memory
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
Persistent DSpace Link
DOI of Published Version
https://doi.org/10.1088/1741-2560/12/4/046004