Ultrasonic Imaging Transceiver Design for CMUT: A Three-Level 30-Vpp Pulse-Shaping Pulser With Improved Efficiency and a Noise-Optimized Receiver

Thumbnail Image
Download
Name

!FinalPaper.pdf

Size

3.34 MB

Format

Adobe PDF

Checksum (MD5)

93542defc7c214682574c9e5d306e3c9

Author(s)
Chen, Kailiang
•
Lee, Hae-Seung
•
Chandrakasan, Anantha P.
•
Sodini, Charles G.
Date Issued
August 2013
Journal
IEEE Journal of Solid-State Circuits
Publisher
Institute of Electrical and Electronics Engineers (IEEE)
Citation
Chen, Kailiang, Hae-Seung Lee, Anantha P. Chandrakasan, and Charles G. Sodini. “Ultrasonic Imaging Transceiver Design for CMUT: A Three-Level 30-Vpp Pulse-Shaping Pulser With Improved Efficiency and a Noise-Optimized Receiver.” IEEE Journal of Solid-State Circuits 48, no. 11 (November 2013): 2734–2745.
Version
Author's final manuscript
Abstract
This paper demonstrates a four-channel transceiver chip for medical ultrasonic imaging, interfacing to the capacitive micromachined ultrasonic transducers (CMUTs). The high-voltage transmitter (Tx) uses a three-level pulse-shaping technique with charge recycling to improve the power efficiency. The design requires minimum off-chip components and is scalable for more channels. The receiver is implemented with a transimpedance amplifier (TIA) topology and is optimized for tradeoffs between noise, bandwidth, and power dissipation. The test chip is characterized with both acoustic and electrical measurements. Comparing the three-level pulser against traditional two-level pulsers, the measured Tx efficiency shows 56%, 50%, and 43% more acoustic power delivery with the same total power dissipation at 2.5, 3.3, and 5.0 MHz, respectively. The CMUT receiver achieves the lowest noise efficiency factor compared with that of the literature (2.1 compared to a previously reported lowest of 3.6, in units of mPA ·√(mW/Hz). In addition, the transceiver chip is tested as a complete system for medical ultrasound imaging applications, in experiments including Tx beamformation, pulse-echo channel response characterization, and ultrasonic Doppler flow rate detection.
MIT Department
Massachusetts Institute of Technology. Department of Electrical Engineering and Computer Science
Terms of Use
Creative Commons Attribution-Noncommercial-Share Alike
http://creativecommons.org/licenses/by-nc-sa/4.0/
Persistent DSpace Link
http://hdl.handle.net/1721.1/93893
DOI of Published Version
https://doi.org/10.1109/jssc.2013.2274895
Repository logo
PrivacyPermissionsAccessibilityContact us
Repository logo
Notify us about copyright concerns.